Unomedical, Inc.

Country not established

Devices
3
FDA recalls
46

At a glance

Counted by deviCERTS from registry records

Market reach

United States
FDA recalls

Catalogue3

Newest first, rows 1-3 of 3

Company details

DUNS number
030271837

FDA recalls46

Recalls are matched to the firm by name. FDA recall data as of 22 Oct 2026.

Showing the 10 most recent of 46 recalls.

  • 2015

    • 21 May 2015 Z-1821-2015 Class II Terminated Unomedical has found that in rare cases the steel needle can break during use, interrupting the delivery of medication including insulin. Distribution: Worldwide Distribution -- US, including the states of CA, IL, MN, and PA; and, the countries of Australia, Bahrain,…
      Full distribution text Worldwide Distribution -- US, including the states of CA, IL, MN, and PA; and, the countries of Australia, Bahrain, Belgium, Canada, China, Cypress, Denmark, Estonia, France, Germany, Greece, Hong Kong, Hungary, Israel, Italy, Kuwait, Lebanon, Malaysia, Maldives, Oman, Schweiz, South Africa, Spain, Sweden, Switzerland, Netherlands, Turkey, United Arab Emirates, and United Kingdom.
      Submission number: K120395
  • 2014

    • 7 Nov 2014 Z-0355-2017 Class II Terminated Tubing detachment Distribution: Worldwide Distribution - US including CA, IN, PA, and Internationally to Canada, Korea, and Japan. Submission numbers: K120872, K972135
    • 7 Nov 2014 Z-0356-2017 Class II Terminated Tubing detachment Distribution: Worldwide Distribution - US including CA, IN, PA, and Internationally to Canada, Korea, and Japan. Submission number: K041545
    • 7 Nov 2014 Z-0357-2017 Class II Terminated Tubing detachment Distribution: Worldwide Distribution - US including CA, IN, PA, and Internationally to Canada, Korea, and Japan. Submission numbers: K120872, K972135
    • 7 Nov 2014 Z-0358-2017 Class II Terminated Tubing detachment Distribution: Worldwide Distribution - US including CA, IN, PA, and Internationally to Canada, Korea, and Japan. Submission numbers: K120872, K972135
    • 7 Nov 2014 Z-0359-2017 Class II Terminated Tubing detachment Distribution: Worldwide Distribution - US including CA, IN, PA, and Internationally to Canada, Korea, and Japan. Submission numbers: K002138, K972135
    • 7 Nov 2014 Z-0360-2017 Class II Terminated Tubing detachment Distribution: Worldwide Distribution - US including CA, IN, PA, and Internationally to Canada, Korea, and Japan. Submission number: K041545
  • 2009

    • 6 Oct 2009 Z-0039-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0040-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0041-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
Show the 25 most recent of 46 recalls
  • 2009

    • 6 Oct 2009 Z-0042-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0043-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0044-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0045-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0046-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0047-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0048-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0049-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0050-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0051-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0052-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0053-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0054-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0055-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286
    • 6 Oct 2009 Z-0056-2010 Unclassified (pre-2012) Terminated In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient. Distribution: Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru,…
      Full distribution text Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
      Submission number: K873286

Frequently asked questions

How many devices does UNOMEDICAL, INC. have in the registry?

The registry currently lists 3 devices for UNOMEDICAL, INC.. Devices recorded under merged duplicate company entries are counted once.

Which registries list UNOMEDICAL, INC.'s devices?

Among its 3 most recently registered devices, UNOMEDICAL, INC. has devices listed in FDA GUDID (3 devices).

What risk classes do UNOMEDICAL, INC.'s devices fall under?

Among its 3 most recently registered devices, the classes recorded for UNOMEDICAL, INC.'s devices are Class II. Each device page shows its own class and authorisation.

Are there FDA recall records for UNOMEDICAL, INC.?

FDA recall records list 46 recalls initiated by UNOMEDICAL, INC., the most recent initiated on 21 May 2015. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.

What are the newest devices from UNOMEDICAL, INC. in the registry?

The most recently registered devices listed for UNOMEDICAL, INC. include Hospitak, Mouth to Mask Resuscitator with Filter Infant Mask 1055-E, Hospitak, Mouth to Mask Resuscitator with Filter Adult Large Mask 1059-E and Hospitak, Mouth to Mask Resuscitator with Filter Child Mask 1056-E.