Vital Images, Inc.
United StatesFDA 510(k) applicant, unverified
- Devices
- 41
- FDA recalls
- 10
At a glance
Market reach
- United States
- FDA recalls
What Vital Images, Inc. makes
Vital Images, Inc. develops medical imaging software and systems for processing and viewing diagnostic scans, including CT, MRI, and dual-energy imaging. Their products support advanced visualization tools such as Vitrea View, Multi Modality Viewer, and specialized software versions like VITREA2 and Vitrea Core. The portfolio also includes AI-assisted analysis tools, such as the V SCORE with Auto Gate model.
The company’s devices are classified as Class II under FDA regulations and are listed via 510(k) clearance. These products are intended for use in the United States and are governed by U.S. medical device compliance standards.
Catalogue
Search and filter all 41 devicesFDA recalls
FDA recall records list 10 recalls initiated by Vital Images, Inc., the earliest on 2007-11-08 and the most recent on 2013-10-21. By recorded status: 10 terminated. The reasons filed fall into these groups: software or firmware and manufacturing or quality process. A recall is often a precaution started by the manufacturer, and a count of records is not a verdict on a company.
2013
- 21 Oct 2013 Z-0169-2014 Class II Terminated Vital Images, Inc. has found an issue in VitreaCore software though internal testing. There is a potential for incorrect and possibly reversed image orientation for certain snapshots taken in VitreaCore software versions 6.0, 6.1, 6.2, 6.3 (except 6.3.1), 6.4 (except 6.4.4 and 6.4.5), as well as their upgrades, and 6.5.
Full distribution text
- 15 Mar 2013 Z-1144-2013 Class II Terminated Vital Images has made a decision to perform a field correction to the Vitrea CT Multi-Chamber Cardiac Functional Analysis application. Vital Images has identified a defect in which numerical values may be incorrect in the Results Table under the Cardiac Analysis portion of the user interface.
Full distribution text
- 15 Mar 2013 Z-1181-2013 Class II Terminated Vital Images, Inc. (Vital Images) received one report about potential user confusion when viewing a brain perfusion scan on Vitrea CT Brain Perfusion 2D when the scan was performed using a wide coverage shuttle acquisition technique. This may result in patient treatment delays, misdiagnosis or mistreatment.
Full distribution text
- 5 Feb 2013 Z-1002-2013 Class II Terminated Vital Images has found a potential error in measurement generated through the software which affects all versions of Vitrea Enterprise Suite, Vitrea, VitreaAdvanced, VitreaCore (VitalConnect) and Vitrea fX prior to version 6.3.3, except 6.1.6, 6.2.3 and 6.3.
This error occurs when images are rotated at the time of reconstruction by the scanner in non-90 degree increments of the transverse/axia
Full distribution text
- 21 Oct 2013 Z-0169-2014 Class II Terminated Vital Images, Inc. has found an issue in VitreaCore software though internal testing. There is a potential for incorrect and possibly reversed image orientation for certain snapshots taken in VitreaCore software versions 6.0, 6.1, 6.2, 6.3 (except 6.3.1), 6.4 (except 6.4.4 and 6.4.5), as well as their upgrades, and 6.5.
2012
- 13 Aug 2012 Z-0186-2013 Class II Terminated Vital Images has found two potential errors in calcium score values in restored snapshots of studies generated through use of Vitrea or Vitrea Enterprise Suite. This issue affects Version 3.1 of Vitrea fX; Version 5.2 of Vitrea; Versions 6.0, 6.1 and 6.2 of Vitrea and Vitrea fX; and Versions 1.3, 6.0, 6.1 and 6.2 of Vitrea Enterprise Suite; and all updates to those versions.
Full distribution text
- 13 Aug 2012 Z-0186-2013 Class II Terminated Vital Images has found two potential errors in calcium score values in restored snapshots of studies generated through use of Vitrea or Vitrea Enterprise Suite. This issue affects Version 3.1 of Vitrea fX; Version 5.2 of Vitrea; Versions 6.0, 6.1 and 6.2 of Vitrea and Vitrea fX; and Versions 1.3, 6.0, 6.1 and 6.2 of Vitrea Enterprise Suite; and all updates to those versions.
2011
- 11 Nov 2011 Z-0388-2012 Unclassified (pre-2012) Terminated Vital Images has found two issues that may display incorrect Standard Update Values (SUV) when viewing certain PET/CT images. This issue affects the VitreaCore PET/CT feature for Vitrea Enterprise Suite 1.2, 1.3, 6.0, 6.1, including all updates to those versions.
Full distribution text
- 11 Nov 2011 Z-0388-2012 Unclassified (pre-2012) Terminated Vital Images has found two issues that may display incorrect Standard Update Values (SUV) when viewing certain PET/CT images. This issue affects the VitreaCore PET/CT feature for Vitrea Enterprise Suite 1.2, 1.3, 6.0, 6.1, including all updates to those versions.
2010
- 14 Oct 2010 Z-0797-2011 Unclassified (pre-2012) Terminated Vital Images Inc., reported that a software defect exists with Vitrea fX 2.1, 3.0 and 3.1 CT Brain Analysis whereby the annotated numeric ROI values on a batched perfusion map are incorrect when created using the auto batching function (from the right click menu) with the auto ROI templates enabled.
Full distribution text
- 23 Aug 2010 Z-1021-2011 Unclassified (pre-2012) Terminated There is a potential alignment error of fused PET/CT images in the Vitrea Core component of Vitrea Enterprise Suite 1.2 and 1.3. The error occurs during panning of the fused PET/CT images. Because the PET data has a larger voxel size than the CT data, it is possible to pan the CT image in more frequent, smaller steps than the PET image, so that the fused PET image does not move until the CT data
Full distribution text
- 23 Feb 2010 Z-1600-2010 Unclassified (pre-2012) Terminated Vital images, Inc is issuing a recall notice on their medical device software in the use of CT Lung and CT Colon in Vitrea 5.1, Vitrea fX 3.0 or Vitre Enterprise Suite 1.2. because incorrect measurements may result. Questions or comments are directed to the firm's Customer Support at [email protected] or at 1-800-208-3005.
Full distribution text
- 14 Oct 2010 Z-0797-2011 Unclassified (pre-2012) Terminated Vital Images Inc., reported that a software defect exists with Vitrea fX 2.1, 3.0 and 3.1 CT Brain Analysis whereby the annotated numeric ROI values on a batched perfusion map are incorrect when created using the auto batching function (from the right click menu) with the auto ROI templates enabled.
2007
- 8 Nov 2007 Z-0520-2008 Unclassified (pre-2012) Terminated Standard Uptake Values (SUV) display/update problems; in the way Standard Uptake Values (SUV) are displayed in 3D/MPR views. The problem arises when a user draws a selected region of interest in order to calculate the SUV statistics of the region. If the user then changes the slice shown by scrolling through additional slices, the original SUV measurements will remain on the screen and do not up
Full distribution text
- 8 Nov 2007 Z-0520-2008 Unclassified (pre-2012) Terminated Standard Uptake Values (SUV) display/update problems; in the way Standard Uptake Values (SUV) are displayed in 3D/MPR views. The problem arises when a user draws a selected region of interest in order to calculate the SUV statistics of the region. If the user then changes the slice shown by scrolling through additional slices, the original SUV measurements will remain on the screen and do not up
Frequently asked questions
How many devices does Vital Images, Inc. have in the registry?
The registry currently lists 41 devices for Vital Images, Inc.. Devices recorded under merged duplicate company entries are counted once.
Which registries list Vital Images, Inc.'s devices?
Among its 41 most recently registered devices, Vital Images, Inc. has devices listed in FDA 510(k) (41 devices).
What risk classes do Vital Images, Inc.'s devices fall under?
Among its 41 most recently registered devices, the classes recorded for Vital Images, Inc.'s devices are Class II. Each device page shows its own class and authorisation.
Are there FDA recall records for Vital Images, Inc.?
FDA recall records list 10 recalls initiated by Vital Images, Inc., the most recent initiated on 2013-10-21. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.
What are the newest devices from Vital Images, Inc. in the registry?
The most recently registered devices listed for Vital Images, Inc. include VITREA2, VERSION 3.4 MEDICAL IMAGE PROCESSING SOFTWARE, Multi Modality Viewer, Vitrea View and VITREA VERSION 4.0.