160MLC Option is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K092145 on 18 Sep 2009, active as of 22 Oct 2026. 11 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

11 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK092145 Date18 Sep 2009ClearedStatusActive

Official records

Recalls

11 recalls · FDA

    • Terminated Z-1493-2012 FDA Not yet classified initiated 3 Apr 2012

    Siemens became aware of a potential safety risk related to the behavior of the 160 MLC when using the "Motion Stop" button followed by "Reset" several times during the treatment of one beam or one segment (issue 1). Additionally, there may be a potential for collision if the operator does not ensure clearance of the extended flat panel positioner for all gantry rotation angles (issue 2).

    • Terminated Z-2404-2011 FDA Unclassified (pre-2012) initiated 15 Apr 2011

    Siemens became aware of a potential safety risk related to the behavior of the 160 MLC when using the "Motion Stop" button followed by "Reset" several times during the treatment of one beam or one segment.

  • 9 recalls, same reason

    • Terminated Z-1039-2013 FDA Class II initiated 5 Mar 13
    • Terminated Z-1037-2013 FDA Class II initiated 5 Mar 13
    • Terminated Z-1036-2013 FDA Class II initiated 5 Mar 13
    • Terminated Z-1040-2013 FDA Class II initiated 5 Mar 13
    • Terminated Z-1043-2013 FDA Class II initiated 5 Mar 13
    • Terminated Z-1042-2013 FDA Class II initiated 5 Mar 13
    • Terminated Z-1044-2013 FDA Class II initiated 5 Mar 13
    • Terminated Z-1041-2013 FDA Class II initiated 5 Mar 13
    • Terminated Z-1038-2013 FDA Class II initiated 5 Mar 13

    Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.

Timeline

12 dated events

  1. Recall initiated FDA recalls Z-1036-2013
  2. Recall initiated FDA recalls Z-1040-2013
  3. Recall initiated FDA recalls Z-1043-2013
  4. Recall initiated FDA recalls Z-1042-2013
  5. Recall initiated FDA recalls Z-1044-2013
  6. Recall initiated FDA recalls Z-1041-2013
  7. Recall initiated FDA recalls Z-1038-2013
  8. Cleared FDA 510(k) K092145
  9. Recall initiated FDA recalls Z-2404-2011
  10. Recall initiated FDA recalls Z-1493-2012

2 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
IYE
Device class
Class II
Regulation
21 CFR Part 892.5050
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code IYE (Accelerator, Linear, Medical).

Other devices from Siemens Medical Solutions USA, Inc.

See all devices from Siemens Medical Solutions USA, Inc. →