Artiste Mv SA
by Siemens Medical Solutions USA, Inc.
510(k) K072485
Artiste Mv SA is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K072485 on 27 Dec 2007, active as of 22 Oct 2026. 24 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK072485 | Date27 Dec 2007 | StatusActive |
Official records
- FDA 510(k) database K072485
Recalls
-
- Terminated Z-1677-2020
Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.
-
- Terminated Z-1204-2019
Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the Auto Field Sequence (AFS) Automatic Motion Protection (AMP) function implemented at the control of the Digital Linear Accelerator.
-
- Terminated Z-1488-2017
Software update
-
- Terminated Z-1212-2017
Siemens became aware of Incorrect values for the rotational collimator position displayed due to mechanical problems of the sensor slide. In addition, there are improper weld seam at the overhead suspension.
-
- Terminated Z-1059-2015
A hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal damage and smoke emission.
-
- Terminated Z-0980-2015
Use of any unauthorized third-party components on Siemens LINACs may lead to patient or operator injury, damage to the LINAC, and can potentially compromise customer warranty. The dark current dose must be considered in the overall dose calculation, or an overdose and serious injury may occur.
-
- Terminated Z-2180-2014
Siemens became aware of a potential issue with the syngo RT Therapist 4.3.SP1 and syngo RT Oncologist 4.3SP1 software applications running on Siemens brand LINAC systems that may experience sporadic system crashes that will cause the inability to unload patient data.
-
- Terminated Z-2174-2014
A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to lead to a collision with the patient.
-
- Terminated Z-1645-2014
Table may lose calibration during patient treatment, which may cause the possibility of mistreatment.
-
- Terminated Z-1652-2014
Siemens Radiation Oncology became aware that customers may be using the Siemens Linear Accelerator in combination with stereotactic accessories which have not been validated as being compatible with Siemens LINAC models.
-
- Terminated Z-0438-2014
A safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there are many long periods of field shape changes in a treatment plan, the dose due to dark current radiation may sum up to a significant value.
-
- Terminated Z-1092-2012
Software upgrade to installed base to the latest release of the RT Therapist version 4.1.144 which corrects multiple safety issues that could lead to patient mistreatment.
-
- Terminated Z-3003-2011
A potential safety issue was discovered when a pause for an imaging segment is inserted (where an imaging segment does not exist) with the ARTISTE and RT Therapist 4.1 system will continue to the next image.
-
- Terminated Z-1422-2011
The firm become aware of a potential issue with ARTISTE system that may result in patient injury should the positioner belt break, and the cam disc (secondary safety mechanism) is not installed and if the flat panel is over the patient.
-
- Terminated Z-1523-2011
A software problem that has the potential for data loss during the transfer of treatment records to the OIS, which may not be recorded in MOSAIQ, and subsequently may lead to mistreatment.
-
- Terminated Z-0449-2011
The robotic lift armature shaft may break and allow the Flat Panel Positioner with the digital flat panel imager to fall into the Flat Panel cradle. If the positioner is over a patient, there is a potential for injury.
-
- Terminated Z-0121-2010
A failed data transfer is possible that may result in data not being recorded in the interfaced system. Siemens Medical Solutions, Inc. USA became aware of this potential issue with its ARTISTE Linear Accelerator's Syngo RT Therapist v4.1 Workspace (containing the Siemens brand of Verify and Record application called PRIMEVIEW) and the LANTIS Oncology Information System product.
-
- Terminated Z-1970-2011
When editing the note type of a previously entered note, then reverting to the original type, data may be lost.
-
- Terminated Z-1064-2011
Potential for Mistreatment-- when treatment plans intended for non-ARTISTE machinese are loaded, if the plans contain unspecified beam parameters, attributes or unknown block codes, the therapy system may remove those parameters.
-
- Terminated Z-1888-2009
Unexpected treatment plan alteration-- Software issue may result in unintended modifications to treatment
-
- Terminated Z-0949-2009
The range of tolerance is to high for the accessory holder, therefore, Wedges and other accessories do not latch to the locking mechanism correctly.
-
- Terminated Z-0857-2009
Under certain conditions, the flat panel may move unexpectedly and collide with the patient.
-
- Terminated Z-0927-2009
If a plan contains a sequence of beams, including at least 1 ARC treatment beam and if during the setup stage of a beam following an ARC beam in the delivery sequence a termination occurs, and the user selects the immediate resumption feature, syngo RT Therapist Express systems will display and record a second delivery of a prior beam and not the delivery of the current beam in certain circumstanc
-
- Terminated Z-0116-2009
Three issues: unexpected rotation of gantry, unexpected movement of table between beams during patient setup, and single exposure images will be overwritten by double exposure images.
Timeline
- Recall initiated FDA recalls Z-1645-2014
- Recall initiated FDA recalls Z-1652-2014
- Recall initiated FDA recalls Z-2174-2014
- Recall initiated FDA recalls Z-2180-2014
- Recall initiated FDA recalls Z-0980-2015
- Recall initiated FDA recalls Z-1059-2015
- Recall initiated FDA recalls Z-1212-2017
- Recall initiated FDA recalls Z-1488-2017
- Recall initiated FDA recalls Z-1204-2019
- Recall initiated FDA recalls Z-1677-2020
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- IYE
- Device class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code IYE (Accelerator, Linear, Medical).
- BRAINLAB MASK SYSTEM by Brainlab Med. Computersysteme GmbH
- WingSTEP Elevation Block ABS P10107-452 by IT-V Medizintechnik GmbH
- Radixact Treatment Delivery System by Accuray Incorporated
- NA TANDEM T10015 by PTW-Freiburg Physikalisch-Technische Werkstätten Dr. Pychlau GmbH
- RefleXion Medical Radiotherapy System by REFLEXION MEDICAL, INC.
- COMPUTER ASSIST STEREOTACTIC SURG SYST, MODIFIED by Medical Instrumentation & Diagnostics Corp.(Midco)
- INTRA OPERATIVE RADIATION THERAPY by Arplay Medical S.A.
- ExacTrac Dynamic by Brainlab AG
- ACCULOC BREAST LOCALIZATION SYSTEM by Northwest Medical Physics Equipment, Inc.
- KLARITY MEDICAL R7504-17.5C by Klarity Medical Products, LLC