5C1 Transducer 11268278
by Siemens Medical Solutions USA, Inc.
Ref 11268278Model MDR_07_IIA_IQH_BTS LB A_ACDV_5C1_11268278
5C1 Transducer 11268278 is an extracorporeal ultrasound imaging transducer (EU MDR Class IIa; FDA Class II; FDA product code ITX, Transducer, ultrasonic, diagnostic) made by Siemens Medical Solutions USA, Inc., registered in EUDAMED on 8 Sep 2026 (Basic UDI-DI 0405686902415VJ), cleared under FDA 510(k) K201130 on 26 May 2020 (1 earlier submission) and listed in FDA GUDID since 20 Feb 2018 (UDI-DI 04056869051932), active as of 23 Oct 2026.
Active
Valid until
No date given
FDA recalls
3 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number0405686902415VJ 04056869051932 | Date8 Sep 2026 | StatusActive |
| United States | FDA 510(k)SESE GUDID | NumberK201130 | Date26 May 2020 | StatusActive |
| United States | GUDID listing GUDID | Number04056869051932 | Date20 Feb 2018 | StatusListed |
Official records
- EUDAMED public site 0405686902415VJ
- AccessGUDID record 04056869051932
- FDA 510(k) database K201130
- FDA 510(k) database K180039
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI 0405686902415VJ registered 8 Sep 2026 Active
- FDA 510(k) K201130 decided 26 May 2020 Active
- FDA GUDID UDI-DI 04056869051932 listed 20 Feb 2018 Listed
Recalls
-
- Ongoing Z-3215-2024
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
-
- Ongoing Z-1599-2024
On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
-
- Ongoing Z-0748-2022
The clip store function in the ultrasound imaging system does not work when the system has a disk full error. This could cause a delay in treatment if the ultrasound system is unable to save clips as study documentation during a high risk procedure, such as a stress echo exam.
-
- Terminated Z-2471-2021
Due to intermittent failures of the power supply in the ultrasound system which renders it inoperable.
Timeline
- Cleared FDA 510(k) K180039
- Listed FDA GUDID 04056869051932
- Cleared FDA 510(k) K201130
- Recall initiated FDA recalls Z-2471-2021
- Recall initiated FDA recalls Z-0748-2022
- Recall initiated FDA recalls Z-1599-2024
- Recall initiated FDA recalls Z-3215-2024
- Registered EUDAMED 0405686902415VJ
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- 5C1 Transducer
- Generic name
- Extracorporeal ultrasound imaging transducer, hand-held
- Device name
- MDR_07_IIA_IQH_BTS Lb A_ACDV_5C1_11268278
- Model / reference
- 11268278, MDR_07_IIA_IQH_BTS LB A_ACDV_5C1_11268278
- Primary DI / UDI-DI
- 04056869051932
- Basic UDI-DI
- 0405686902415VJ
- FDA product code
- ITX
- EMDN code
- Z110402010201 Curvilinear transcutaneous two–dimensional ultrasound probes (convex–array)
- GMDN term
- Extracorporeal ultrasound imaging transducer, hand-held
- Device class
- Class IIa EUDAMEDClass II GUDID
- Regulation number
- 892.1570
- EU registration type
- Standard device
- EU market of registration
- Germany
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class IIa | Class II |
Similar devices in this category
Other devices filed under CURVILINEAR TRANSCUTANEOUS TWO–DIMENSIONAL ULTRASOUND PROBES (CONVEX–ARRAY).
- C2-6b-D A7MP by KONICA MINOLTA, INC.
- C60XI/5-2 MHZ P20402-11A by Fujifilm Sonosite, Inc.
- C251 04573596211335 by FUJIFILM CORPORATION
- C1-5-RS H40462LA by GE Healthcare Japan Corporation
- C60xf/5-2 MHz P18024-22A by Fujifilm Sonosite, Inc.
- SC3123 122000500 by Esaote S.p.A.
- CA123 9600158000 by Esaote S.p.A.
- C23 04573596211304 by FUJIFILM CORPORATION
- C1-5-D H40452LE by GE Healthcare Japan Corporation
- 2D Abdominal Probe K2DAB00 by KOELIS