8F FlowGate2 Balloon Guide Catheter

by Concentric Medical

510(k) K153729

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK153729 Date27 Jan 2016ClearedStatusActive

Recalls

2 recalls · 1 ongoing · FDA

    • Ongoing Z-1907-2024 FDA Class II initiated 23 Apr 2024

    Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

    • Terminated Z-2569-2018 FDA Class II initiated 11 Apr 2018

    There may be product manufactured with the outer diameter of the balloon catheter exceeding the defined specification at a customer location. If the user experiences resistance between the FlowGate and an interfacing device, the device would be removed and a second device would be used, which could potentially prolong a procedure.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DQY
Device Class
Class II
Regulation
21 CFR Part 870.1250
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Other devices from Concentric Medical, Inc.

See all devices from Concentric Medical, Inc. →