ABL800 Flex with Flexq Module, Model ABL8XX
510(k) K043218
ABL800 Flex with Flexq Module, Model ABL8XX is a medical device (FDA Class II; FDA product code CHL, Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph) made by Radiometer Medical ApS, cleared under FDA 510(k) K043218 on 10 May 2005, active as of 11 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK043218 | Date10 May 2005 | StatusActive |
Official records
- FDA 510(k) database K043218
From the FDA record
- Received by the FDA
- 22 Nov 2004
- FDA decision
- 10 May 2005 (Substantially Equivalent)
- FDA review time
- 169 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-2303-2013
RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.
-
- Terminated Z-2571-2011
Based on internal testing and user/customer complaint reports received, the firm determined that wrinkles may arise on the outer membrane of the calcium membrane units upon installation. Retention of fluids between the wrinkles can cause a positive bias resulting in calcium measuring errors. This situation poses a risk to the safety of patients, particularly when the measurement results are used
-
- Terminated Z-0705-2007
In some situations pre-registered patient data is not reset when leaving the pre-registration mode in the software. This occurs when a Safe Pico blood sampler associated with a specific patient is pre-registered in the analyzer by scanning a barcode.
Timeline
- Cleared FDA 510(k) K043218
- Recall initiated FDA recalls Z-0705-2007
- Recall initiated FDA recalls Z-2571-2011
- Recall initiated FDA recalls Z-2303-2013
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- CHL
- Device class
- Class II
- Regulation
- 21 CFR Part 862.1120
- Advisory committee
- Clinical Chemistry
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code CHL (Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph).
- ABL 77 PH, Blood Gas and Electrolyte Analysis System by Sendx Medical, Inc.
- ABL510 Blood Gas System by RADIOMETER AMERICA, INC.
- Life Watch Respiratory Monitor by The Perkin-Elmer Corp.
- OMNI MODULAR, OMNI C AND OMNI S ANALYZER (PH ELECTRODE) by Roche Diagnostics Corp.
- Analyzer Nova Stat Profile 2/QUALITY Control Mat by NOVA BIOMEDICAL CORPORATION
- RAPIDLab® 1200 Series 1240 10321840 5060298 by Siemens Healthcare Diagnostics Inc.
- GEM Premier 3000 iQM - BG HCT ELECT. GL - 75 Tests 00024307589 by INSTRUMENTATION LABORATORY COMPANY
- ELECTROLYLE CARBON DIOXIDE ELECTRODES by Fisher Scientific Co., LLC
- PH 6.838 REF. BUFFER by Fisher Scientific Co., LLC
- Stat Profile pHOx Basic Analyzer by NOVA BIOMEDICAL CORPORATION