ABL8X7 Flex
510(k) K051968
ABL8X7 Flex is a medical device (FDA Class II; FDA product code CGL, Electrode, Ion Based, Enzymatic, Creatinine) made by Radiometer Medical ApS, cleared under FDA 510(k) K051968 on 16 Oct 2006, active as of 10 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK051968 | Date16 Oct 2006 | StatusActive |
Official records
- FDA 510(k) database K051968
From the FDA record
- Received by the FDA
- 20 Jul 2005
- FDA decision
- 16 Oct 2006 (Substantially Equivalent)
- FDA review time
- 453 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-0801-2019
From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on analyzers could be underestimated with up to 28% as compared with measurements from COBAS. The underestimated creatinine occurs specifically, when measurements on Autocheck 6+ level 1 (also known as S7835 Autocheck module and abbreviated AC6+L1) are less than or equal to 230 micro-meter of a specified insert interval for AC6+L1 with insert limits 210 290 micro-meter. Ultimately, a lower read-out on patient samples is observed Customers are concerned about underestimation of creatinine results in patients, which are to undergo radiocontrast x-ray investigations. If these patients are not healthy, the exposure of radiocontrast agents may induce kidney insufficiency (contrast-induced nephropathy). Hazardous situation: Underestimation of the concentration of creatinine.
-
- Terminated Z-1293-2015
Negative drift on QC during in-use lifetime of CREA membranes can be observed.
-
- Terminated Z-1008-2012
When the user activates the dosicap (containing the crea additive) and mounts it onto the bottle, the threads will not seal correctly causing approximately 10% of the solution to leak out when the bottle is placed horizontally for dissolving the additive. This will cause the concentration of creatinin and creatin in the calibration solution to be wrong, and results in a bias on the Crea results.
-
- Terminated Z-1015-2010
Analysis of customer data, (for ABL 8X7 customers) has shown that the reference membranes are out of analytical specifications when replaced as specified in the device labeling. These errors in analytical specifications could lead to inadequate or clinical interventions (therapy) or misleading interpretation of patient status in clinical settings.
Timeline
- Cleared FDA 510(k) K051968
- Recall initiated FDA recalls Z-1015-2010
- Recall initiated FDA recalls Z-1008-2012
- Recall initiated FDA recalls Z-1293-2015
- Recall initiated FDA recalls Z-0801-2019
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- CGL
- Device class
- Class II
- Regulation
- 21 CFR Part 862.1225
- Advisory committee
- Clinical Chemistry
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code CGL (Electrode, Ion Based, Enzymatic, Creatinine).
- i-STAT 03P84-25 by Abbott Point of Care Inc.
- i-STAT Crea cartridge with the i-STAT 1 System by Abbott Point of Care Inc.
- NOVA STATSENSOR CREATININE HOSPITAL METER, CONTROL SOLUTIONS, LINEARITY SOLUTIONS, AND TEST STRIPS by NOVA BIOMEDICAL CORPORATION
- I-STAT SYSTEM by I-Stat Corp.
- Nova StatSensor Creatinine Hospital Meter System by NOVA BIOMEDICAL CORPORATION
- RENALYZER PRX 90 by Swedish Trade Council USA
- EPOC CHLORIDE TEST AND EPOC CREATININE TEST by Epocal Inc