ABL90 Flex Plus
510(k) K160153
ABL90 Flex Plus is a medical device (FDA Class II; FDA product code CHL, Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph) made by Radiometer Medical ApS, cleared under FDA 510(k) K160153 on 4 Nov 2016, active as of 11 Oct 2026. 4 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK160153 | Date4 Nov 2016 | StatusActive |
Official records
- FDA 510(k) database K160153
From the FDA record
- Received by the FDA
- 22 Jan 2016
- FDA decision
- 4 Nov 2016 (Substantially Equivalent)
- FDA review time
- 287 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-1659-2025
ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
-
- Ongoing Z-1858-2021
- Ongoing Z-1857-2021
Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed.
-
- Terminated Z-1761-2020
The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.
-
- Terminated Z-2138-2020
The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.
Timeline
- Cleared FDA 510(k) K160153
- Recall initiated FDA recalls Z-2138-2020
- Recall initiated FDA recalls Z-1761-2020
- Recall initiated FDA recalls Z-1858-2021
- Recall initiated FDA recalls Z-1857-2021
- Recall initiated FDA recalls Z-1659-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- CHL
- Device class
- Class II
- Regulation
- 21 CFR Part 862.1120
- Advisory committee
- Clinical Chemistry
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code CHL (Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph).
- BLOOD GAS ANALYSIS, STANDARDS CONTROLS by United Diagnostics, Inc.
- IRMA BLOOD ANALYSIS SYSTEM by Diametrics Medical, Ltd.
- GEM Premier 3000 iQM - BG HCT ELECT. GL - 600 Tests 00024360089 by INSTRUMENTATION LABORATORY COMPANY
- BLOOD GASES/ PO2, PCO2 AND BLOOD PH ANALYSIS PROD. by D.R.G. Scientific Industries
- EASYBLOOD GAS ANALYZER by Medica Corp.
- i-STAT CG4+ cartridge with the i-STAT 1 System by Abbott Point of Care Inc.
- RAPIDLab® 1200 Series 1265 10321852 5063769 by Siemens Healthcare Diagnostics Inc.
- GEM Premier 3000 iQM - BG HCT ELECT. GL - 300 Tests 00024330089 by INSTRUMENTATION LABORATORY COMPANY
- STAT PROFILE PHOX PLUS L ANALYZER by NOVA BIOMEDICAL CORPORATION
- i-STAT 03P78-25 by Abbott Point of Care Inc.