Abx Pentra Reference 1220001719
by Horiba Abx
Ref A11A01719
Abx Pentra Reference 1220001719 is a device classified as IVD General (EU), Class II (US) made by Horiba Abx, registered in EUDAMED on 16 Jun 2026 (Basic UDI-DI B-03610230005446), cleared under FDA 510(k) K052007 on 16 Dec 2005 and listed in FDA GUDID since 16 Aug 2016 (UDI-DI 03610230005446), active as of 25 Oct 2026.
Active
Valid until
No date given
FDA recalls
None ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/IVDD EUDAMED | NumberB-03610230005446 03610230005446 | Date16 Jun 2026 | StatusActive |
| United States | GUDID listing GUDID | Number03610230005446 | Date16 Aug 2016 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK052007 | Date16 Dec 2005 | StatusActive |
Official records
- EUDAMED public site B-03610230005446
- AccessGUDID record 03610230005446
- FDA 510(k) database K052007
Across registries
- EUDAMED Basic UDI-DI B-03610230005446 registered 16 Jun 2026 Active
- FDA 510(k) K052007 decided 16 Dec 2005 Active
- FDA GUDID UDI-DI 03610230005446 listed 16 Aug 2016 Listed
Certificates
No certifier certificate is listed. Devices of this class are self-declared by the manufacturer.
Recalls
-
- Terminated Z-0493-2018
- Terminated Z-0494-2018
The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
-
- Terminated Z-0740-2016
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
-
- Terminated Z-0902-2015
HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. The defect would result in a false high or false low results for HbA1c and false high results for magnes
-
- Terminated Z-0904-2014
Horiba Medical is recalling the ABX PENTRA N Control because it has been updated to include a revised Target Value assignment for Potassium.
-
- Terminated Z-1141-2013
HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.
Timeline
- Cleared FDA 510(k) K052007
- Recall initiated FDA recalls Z-1141-2013
- Recall initiated FDA recalls Z-0904-2014
- Recall initiated FDA recalls Z-0902-2015
- Recall initiated FDA recalls Z-0740-2016
- Listed FDA GUDID 03610230005446
- Recall initiated FDA recalls Z-0493-2018
- Recall initiated FDA recalls Z-0494-2018
- Registered EUDAMED B-03610230005446
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Abx Pentra Reference
- Generic name
- Multiple blood gas/haemoximetry/electrolyte analyte IVD, calibrator
- Model / reference
- A11A01719 EUDAMED1220001719 GUDID
- Primary DI / UDI-DI
- 03610230005446
- Basic UDI-DI
- B-03610230005446
- 510(k) Number
- FDA product code
- JIX
- EMDN code
- W0101050301 Calibrators multicomponent (cc)
- GMDN term
- Multiple blood gas/haemoximetry/electrolyte analyte IVD, calibrator
- Device class
- Class IVD General EUDAMEDClass II GUDID
- Regulation number
- 862.1150
- EU registration type
- Legacy device (under a directive)
- EU market of registration
- France
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / reference | A11A01719 | 1220001719 |
| Device class | Class IVD General | Class II |
Similar devices in this category
Other devices filed under CALIBRATORS MULTICOMPONENT (CC).
- Atellica CH ENZ 2 Calibrator 11099318 by Siemens Healthcare Diagnostics Inc.
- Plasma Calibrator Set 92025/XT by Chromsystems Instruments & Chemicals GmbH
- Plasma Calibrator Set 75128 by Chromsystems Instruments & Chemicals GmbH
- Dimension Vista® CHEM 4 CAL 10445172 KC140 by Siemens Healthcare Diagnostics Inc.
- ELICAL 2 CALI-0550 by VITALSCIENTIFIC
- ARCHITECT 1E65-06 01E6506 by Abbott
- Wiener Lab. Calibrator A Plus, Model 2X3 mL Cat. Nr. 1918006 by Wiener Laboratories Saic
- MEDI-CAL 1578-0891 by Medicon Hellas SA
- SYNCHRON Systems OP 300 Low Urine Control 474988 by Beckman Coulter, Inc.
- CERULOPLASMIN 13340 by BIOSYSTEMS S.A.