Accelerator a3600
by Inpeco SA
Ref ACPModel TLA
Accelerator a3600 (ACP), a specimen processing instrument IVD (EU IVDR Class A; FDA Class II; FDA product code CEM, Electrode, Ion Specific, Potassium) made by Inpeco SA, was registered in EUDAMED on 23 May 2023 (Basic UDI-DI 764017234TLAKW), cleared under FDA 510(k) K121012 on 31 Aug 2012 and listed in FDA GUDID on 24 Sep 2016 (UDI-DI 07640172341001); the registry lists it as withdrawn (EUDAMED) as of 25 Oct 2026. FDA recalls on record: 24.
Active
Valid until
No expiry
FDA recalls
17 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/IVDR EUDAMED | Number764017234TLAKW 07640172341001 | Date23 May 2023 | StatusWithdrawn |
| United States | GUDID listing GUDID | Number07640172341001 | Date24 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK121012 | Date31 Aug 2012 | StatusActive |
Official records
- EUDAMED public site 764017234TLAKW
- AccessGUDID record 07640172341001
- FDA 510(k) database K121012
From the FDA record
- Device description
- The Accelerator a3600 automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system consolidates multiple Analytical instruments into a unified workstation.
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI 764017234TLAKW registered 23 May 2023 Withdrawn
- FDA 510(k) K121012 decided 31 Aug 2012 Active
- FDA GUDID UDI-DI 07640172341001 listed 24 Sep 2016 Listed
Recalls
-
- Ongoing Z-0595-2026
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.
-
- Ongoing Z-0596-2026
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
-
- Ongoing Z-0107-2026
The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
-
- Ongoing Z-0106-2026
The FlexLab System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
-
- Ongoing Z-0560-2024
Firmware of the Interface Module with Roche Cobas 8000 and Cobas PRO (CO8 IM) may lead to a delayed sample tube processing.
-
- Ongoing Z-1562-2023
- Ongoing Z-1563-2023
- Ongoing Z-1561-2023
- Ongoing Z-1559-2023
- Ongoing Z-1560-2023
Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise. The event may occur only if all the following conditions occurs in few milliseconds timeframe: - The module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B. Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
-
- Ongoing Z-0087-2022
The Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead to an incorrect association of test results to sample ID. Improper management of the sample may bring an incorrect patient result.
-
- Ongoing Z-1889-2020
- Ongoing Z-1888-2020
- Ongoing Z-1887-2020
The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
-
- Ongoing Z-2198-2020
Potential for discrepant patient results due to a potential for the primary sample tube to be diluted. In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.
-
- Ongoing Z-2199-2020
- Ongoing Z-2200-2020
In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.
-
- Terminated Z-2975-2020
- Terminated Z-2974-2020
- Terminated Z-2973-2020
Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
-
- Terminated Z-0432-2021
- Terminated Z-0434-2021
- Terminated Z-0433-2021
When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.
-
- Terminated Z-2115-2020
- Terminated Z-2114-2020
The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.
-
- Terminated Z-2003-2020
- Terminated Z-2004-2020
When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.
-
- Terminated Z-2141-2020
- Terminated Z-2140-2020
- Terminated Z-2142-2020
The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.
-
- Terminated Z-0361-2019
The firm has identified customer sites where there are modules on which the safety pacemaker label is missing. The information regarding the label included in the Operations Manual has been updated to specify that: 1) the risk of a pacemaker malfunction applies to anyone with a pacemaker who may work on these modules; and 2) the minimum safety distance has been increased to 200 mm (7.87") from the transport mechanisms. Any person with a pacemaker must not get closer than this distance from the modules axes.
-
- Terminated Z-1798-2018
Module may freeze without generating user warning. There is a potential risk in delay of sample processing, leading to delayed delivery of test results to patients.
-
- Terminated Z-0587-2017
Inpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets.
Timeline
- Recall initiated FDA recalls Z-1563-2023
- Recall initiated FDA recalls Z-1561-2023
- Recall initiated FDA recalls Z-1559-2023
- Recall initiated FDA recalls Z-1560-2023
- Registered EUDAMED 764017234TLAKW
- Recall initiated FDA recalls Z-0560-2024
- Recall initiated FDA recalls Z-0107-2026
- Recall initiated FDA recalls Z-0106-2026
- Recall initiated FDA recalls Z-0595-2026
- Recall initiated FDA recalls Z-0596-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Generic name
- Specimen processing instrument IVD
- Device name
- Inpeco Total Laboratory Automation
- Model / reference
- TLA, ACP
- Primary DI / UDI-DI
- 07640172341001
- Basic UDI-DI
- 764017234TLAKW
- Direct marketing DI
- 07640172341001
- 510(k) Number
- FDA product code
- CEM
- EMDN code
- W02060101 Automated systems for the management of pre and post-analysis
- GMDN term
- Specimen processing instrument IVD
- Device class
- Class A EUDAMEDClass II GUDID
- Regulation number
- 862.1600
- EU registration type
- Standard device
- EU market of registration
- France
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class A | Class II |
Manufacturer details
- EUDAMED SRN
- CH-MF-000017833
- Authorised representative
- QbD RepS BV BE-AR-000000040
Similar devices in this category
Other devices filed under AUTOMATED SYSTEMS FOR THE MANAGEMENT OF PRE AND POST-ANALYSIS.
- Automatic Whole Rack Cryogenic Tubes Capper/Decapper BM48 by Hangzhou Biobank Bio-technology Co., Ltd.
- Multi-channel hand-held Screw Cap Recapper BQ-OP6 by Hangzhou Biobank Bio-technology Co., Ltd.
- Automatic Sample Processing System XQ-0057-05 by Shenzhen Dymind Biotechnology Co., Ltd.
- Collaborative Station W092 by Copan Wasp S.r.l.
- m2000 09K14-02 by Abbott Molecular
- DxA 5000, DxA 5000 Fit B87352 by Beckman Coulter Biomedical GmbH
- Alinity mp 09N65-001 by Abbott Molecular
- Automated Liquid Handling Workstation HBLH-9600A by Hybribio Medtech Device Co.,Ltd
- UniVerse W093 by Copan Wasp S.r.l.
- Atellica Direct Connect for CI 11767236 by Siemens Healthcare Diagnostics Inc.