Inpeco SA
Active in EUDAMEDSwitzerlandfrom SRN EUDAMED FDA SRN CH-MF-000017833 FEI 3010825766
Inpeco SA is a medical device manufacturer based in Switzerland, with 8 devices listed in EUDAMED and FDA GUDID (EU Class A; FDA Class I) and 33 FDA recalls as of 18 Nov 2026.
- Devices
- 8
- Registrations
- EUDAMEDFDA
- FDA recalls
- 3311 ongoing since 2021 (17 open in total)
Registry facts
- Registries
- EUDAMED 7
- FDA GUDID 7
- Device classes
- EU Class A 7
- FDA Class I 1
- Top EMDN categories
- FDA recalls
- Sites and related companies
- QbD RepS BV authorised representative
- Identifiers
- EUDAMED SRN CH-MF-000017833
- FDA FEI 3010825766
- DUNS 484880690
At a glance
Market reach
- European Union
- EUDAMED actor
- United States
- FDA FEI, FDA recalls
- Authorised rep.
- QbD RepS BVBE
Catalogue
- Regimes: CE IVDR Registry: GUDID
FlexLab Ref FLX
Regimes: CE IVDR Registry: EUDAMEDGUDIDFlexLab X Ref FXX
Regimes: CE IVDR Registry: EUDAMEDGUDIDAptio® Automation Ref AP2
Regimes: CE IVDR Registry: EUDAMEDGUDIDProTube Ref PTB
Regimes: CE IVDR Registry: EUDAMEDGUDIDAccelerator a3600 Ref ACP
Regimes: CE IVDR, withdrawn Registry: EUDAMEDGUDIDFlexPath Blox Ref FPB
Regimes: CE IVDR Registry: EUDAMEDGUDIDFlexPath Blox Ref FPB
Regimes: CE IVDR, withdrawn Registry: EUDAMED
Company details
- EUDAMED SRN
- CH-MF-000017833
- FDA FEI number
- 3010825766
- DUNS number
- 484880690
- EU VAT number
- CHE-114.538.298
- EUDAMED status
- Active
- EUDAMED registered name
- Inpeco SA
- Registered address
- 26, Via Torraccia, 6883, Novazzano, Switzerland
- PRRC contact
- Luisa Mella
FDA recalls
Recalls initiated per year
- 2017 0
- 2018 2
- 2019 0
- 2020 19
- 2021 1
- 2022 0
- 2023 6
- 2024 0
- 2025 4
- 2026 0
2025
- 30 Oct 2025 Z-0595-2026 Class II Ongoing The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.
- 30 Oct 2025 Z-0596-2026 Class II Ongoing The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
- 3 Sep 2025 Z-0106-2026 Class II Ongoing The FlexLab System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
- 3 Sep 2025 Z-0107-2026 Class II Ongoing The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
2023
- 3 Nov 2023 Z-0560-2024 Class II Ongoing Firmware of the Interface Module with Roche Cobas 8000 and Cobas PRO (CO8 IM) may lead to a delayed sample tube processing.
Full distribution text
- 10 Mar 2023 Z-1559-2023 Class II Ongoing Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise.
The event may occur only if all the following conditions occurs in few milliseconds timeframe:
- The module is releasing a sample tube (Tube A) just placed into the carrier
- Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate
Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message.
The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B.
Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
Full distribution text
- 10 Mar 2023 Z-1560-2023 Class II Ongoing Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise.
The event may occur only if all the following conditions occurs in few milliseconds timeframe:
- The module is releasing a sample tube (Tube A) just placed into the carrier
- Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate
Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message.
The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B.
Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
Full distribution text
- 10 Mar 2023 Z-1561-2023 Class II Ongoing Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise.
The event may occur only if all the following conditions occurs in few milliseconds timeframe:
- The module is releasing a sample tube (Tube A) just placed into the carrier
- Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate
Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message.
The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B.
Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
Full distribution text
- 10 Mar 2023 Z-1562-2023 Class II Ongoing Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise.
The event may occur only if all the following conditions occurs in few milliseconds timeframe:
- The module is releasing a sample tube (Tube A) just placed into the carrier
- Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate
Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message.
The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B.
Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
Full distribution text
- 10 Mar 2023 Z-1563-2023 Class II Ongoing Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise.
The event may occur only if all the following conditions occurs in few milliseconds timeframe:
- The module is releasing a sample tube (Tube A) just placed into the carrier
- Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate
Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message.
The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B.
Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
Full distribution text
- 3 Nov 2023 Z-0560-2024 Class II Ongoing Firmware of the Interface Module with Roche Cobas 8000 and Cobas PRO (CO8 IM) may lead to a delayed sample tube processing.
Show the 25 most recent of 33 recalls
2021
- 10 Sep 2021 Z-0087-2022 Class II Ongoing The Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead to an incorrect association of test results to sample ID. Improper management of the sample may bring an incorrect patient result.
2020
- 7 Aug 2020 Z-2973-2020 Class II Terminated Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the
potential to mis-associate sample IDs leading to incorrect results or delayed results
Full distribution text
- 7 Aug 2020 Z-2974-2020 Class II Terminated Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the
potential to mis-associate sample IDs leading to incorrect results or delayed results
Full distribution text
- 7 Aug 2020 Z-2975-2020 Class II Terminated Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the
potential to mis-associate sample IDs leading to incorrect results or delayed results
Full distribution text
- 14 Jul 2020 Z-0432-2021 Class II Terminated When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is
flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.
Full distribution text
- 14 Jul 2020 Z-0433-2021 Class II Terminated When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.
Full distribution text
- 14 Jul 2020 Z-0434-2021 Class II Terminated When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.
Full distribution text
- 7 Apr 2020 Z-1887-2020 Class II Open since 2020 The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Full distribution text
- 7 Apr 2020 Z-1888-2020 Class II Open since 2020 The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Full distribution text
- 7 Apr 2020 Z-1889-2020 Class II Open since 2020 The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Full distribution text
- 7 Apr 2020 Z-2003-2020 Class II Terminated When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.
Full distribution text
- 7 Apr 2020 Z-2004-2020 Class II Terminated When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.
Full distribution text
- 7 Apr 2020 Z-2114-2020 Class II Terminated The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.
- 7 Apr 2020 Z-2115-2020 Class II Terminated The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.
- 7 Apr 2020 Z-2140-2020 Class II Terminated The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.
- 7 Aug 2020 Z-2973-2020 Class II Terminated Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the
potential to mis-associate sample IDs leading to incorrect results or delayed results
Frequently asked questions
How many devices does Inpeco SA have in the registry?
The registry currently lists 8 devices for Inpeco SA. Devices recorded under merged duplicate company entries are counted once.
Which registries list Inpeco SA's devices?
Among its 8 most recently registered devices, Inpeco SA has devices listed in EUDAMED (7 devices) and FDA GUDID (7 devices). A device can be listed in more than one registry, so these counts overlap.
What risk classes do Inpeco SA's devices fall under?
Among its 8 most recently registered devices, the classes recorded for Inpeco SA's devices are Class I and IVDR Class A. Each device page shows its own class and authorisation.
Are there FDA recall records for Inpeco SA?
FDA recall records list 33 recalls initiated by Inpeco SA, the most recent initiated on 30 Oct 2025. 17 of them are recorded as ongoing. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.
What are the newest devices from Inpeco SA in the registry?
The most recently registered devices listed for Inpeco SA include ProTube PTB, FlexLab, FlexLab X and Aptio® Automation.