Acuson Antares Diagnostic Ultrasound System
by Siemens Medical Solutions USA, Inc., Ultrasound DI
510(k) K063803
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Overview
The ACUSON ANTARES DIAGNOSTIC ULTRASOUND SYSTEM is a highly advanced diagnostic ultrasound system designed for radiologists to enhance their diagnostic capabilities. With its exceptional image quality, ease of use, and reliability, this system has become a staple in many medical institutions around the world.
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK063803 | Date5 Jan 2007 | StatusActive |
Recalls
-
- Terminated Z-1836-2011
There is an error in the estimation of mean pressure gradient in the Cardiac measurement package. The currently implemented calculation will result in an underestimation of the mean pressure gradient (PGmean) when estimated from a trace of the Doppler waveform of the stenotic flow.
-
- Terminated Z-1853-2011
Image captured from the device may be erroneous.
-
- Terminated Z-1005-2011
- Terminated Z-1006-2011
- Terminated Z-1007-2011
ECG waveform does not accurately align with 2D, M-Mode or Spectral Images.
-
- Terminated Z-1766-2008
- Terminated Z-1767-2008
Software issue may result in distorted images and inaccurate measurements, which could lead to a misdiagnosis.
-
- Terminated Z-0298-2008
Image error: A software problem results in on-screen indications (graphics) that lead the user to believe that the patient's right and left and the transducer orientation are oriented the same. This is not the case and the actual patient's right and left are reversed from the on-screen indicators.
-
- Terminated Z-0297-2008
Image errors: A software problem results in on-screen indications (graphics) that lead the user to believe that the patient's right and left and the transducer orientation are oriented the same. This is not the case and the actual patient's right and left are reversed from the on-screen indicators.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IYN
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1550
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent