Advisor, Model 9200
by Bci, Inc.
510(k) K030407
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK030407 | Date19 Aug 2003 | StatusActive |
Recalls
-
- Terminated Z-2379-2018
Some printing modes on the devices may not function properly (tabular data printing, real-time continuous printouts, and automatic interrupt print feature).
-
- Terminated Z-1692-2015
The Auxiliary Serial Input/ Output Modes on the Advisor do not function as expected, resulting in data formatting issues or incorrect data being transmitted in or out of the serial port into a data collection system.
-
- Terminated Z-0918-2010
Smiths Medical is conducting a voluntary recall of a limited number of medical device AC power cords, manufactured by Electri-Cord Manufacturing Co., for use with BCI 3180 Pulse Oximeter(Human), 9200 Advisor Vital Signs Monitor(Human or Veterinary), 8400 Capnocheck II Capnometer(Human or Veterinary) and V6400 Invasive Pressure Monitor (Veterinary). The affected Electri-Cord AC power cords are eq
-
- Terminated Z-1326-2009
Recent improvements were made to the Advisor case plastic resulting in the inserts being deeply embedded into the plastic. This affects the ability of the monitor to be adequately secured on the wall mount bracket or roll stand bracket, resulting in the monitor falling off. There have been no adverse events or injuries reported.
-
- Terminated Z-1089-2007
Turn off unexpectedly/ Not turn on : monitors turned off unexpectedly sounding the critical failure alarm. Advisor Vital Signs monitors that were off, would not turn back on.
-
- Terminated Z-0238-06
- Terminated Z-0232-06
- Terminated Z-0233-06
- Terminated Z-0222-06
- Terminated Z-0240-06
- Terminated Z-0230-06
- Terminated Z-0225-06
- Terminated Z-0224-06
- Terminated Z-0234-06
- Terminated Z-0237-06
- Terminated Z-0220-06
- Terminated Z-0236-06
- Terminated Z-0226-06
- Terminated Z-0229-06
- Terminated Z-0239-06
- Terminated Z-0221-06
- Terminated Z-0223-06
- Terminated Z-0231-06
- Terminated Z-0228-06
- Terminated Z-0227-06
- Terminated Z-0235-06
Potential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the primary lead is set to Lead II, III or any of the augmented leads, AVL, AVR or AVF on the Advisor monitor.
-
- Terminated Z-0826-05
- Terminated Z-0832-05
- Terminated Z-0828-05
- Terminated Z-0827-05
- Terminated Z-0835-05
- Terminated Z-0834-05
- Terminated Z-0829-05
- Terminated Z-0830-05
- Terminated Z-0825-05
- Terminated Z-0824-05
- Terminated Z-0833-05
- Terminated Z-0831-05
Altitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard temperature pressure). The % CO values displayed will be lower than the actual CO2 value.
-
- Terminated Z-0531-05
- Terminated Z-0532-05
- Terminated Z-0533-05
- Terminated Z-0535-05
- Terminated Z-0534-05
An issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CCK
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.1400
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent