Aloka Arietta 850
by Hitachi Healthcare Americas Corporation 3 variants
510(k) K183456
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK183456 | Date28 Jan 2019 | StatusActive |
Recalls
-
- Ongoing Z-0190-2022
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
-
- Ongoing Z-2308-2021
The screw cover at the tip of the probe may fall off after a cleaning and sterilization process. Users advised to conduct a preventive inspection of the probe before and after sterilization/disinfection/cleaning and before and after an interoperative procedure.
-
- Terminated Z-1282-2021
Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the C252 probe, this error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IYN
- FDA applicant
- Hitachi Healthcare Americas
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1550
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 19 Sep 2026
The Aloka Arietta 850 is an Ultrasonic Pulsed Doppler Imaging System, designed for real-time visualization of internal body structures and blood flow using high-frequency sound waves. It falls under the System, Imaging, Pulsed Doppler, Ultrasonic category (Product Code: IYN) and incorporates advanced Doppler technology to assess vascular and cardiac function.
This device is intended for use in clinical settings such as hospitals, diagnostic imaging centers, and outpatient facilities. Supplied as a reusable medical system, it requires proper maintenance and calibration per manufacturer guidelines. Authorized under 21 CFR 892.1550 and cleared via FDA 510(k) (K183456), it adheres to Class II – Special Controls regulatory standards for pulsed Doppler ultrasound systems.
Other devices from Hitachi Healthcare Americas Corporation
See all devices from Hitachi Healthcare Americas Corporation →