Anesthesia Machine
by Shenzhen Mindray Bio-Medical Electronics Co., Ltd. 2 variants
Ref A8/A9Model A8/A8C /A8 Pro/A80 /A80C/ A9/A9C /A9 Pro/A90 /A90C
Anesthesia Machine (A8/A9) is an anaesthesia unit (EU MDD Class IIb; FDA Class II; FDA product code BSZ, Gas-Machine, Anesthesia) made by Shenzhen Mindray Bio-Medical Electronics Co., Ltd., registered in EUDAMED on 17 May 2023 (Basic UDI-DI B-06936415912284), cleared under FDA 510(k) K201957 on 26 Mar 2021 and listed in FDA GUDID since 15 Jul 2021 (UDI-DI 06936415912284), active as of 25 Oct 2026. 6 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
6 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDD EUDAMED | NumberB-06936415912284 06936415912284 | Date17 May 2023 | StatusActive |
| United States | GUDID listing GUDID | Number06936415912284 | Date15 Jul 2021 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK201957 | Date26 Mar 2021 | StatusActive |
Official records
- EUDAMED public site B-06936415912284
- AccessGUDID record 06936415912284
- FDA 510(k) database K201957
From the FDA record
- MRI safety
- MR Unsafe
- Natural rubber latex
- Labelled as containing natural rubber latex
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI B-06936415912284 registered 17 May 2023 Active
- FDA 510(k) K201957 decided 26 Mar 2021 Active
- FDA GUDID UDI-DI 06936415912284 listed 15 Jul 2021 Listed
Recalls
-
- Ongoing Z-0720-2026
- Ongoing Z-0721-2026
Potential for anesthesia leakage.
-
- Ongoing Z-2949-2024
- Ongoing Z-2950-2024
- Ongoing Z-2951-2024
- Ongoing Z-2952-2024
V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
Timeline
- Cleared FDA 510(k) K201957
- Listed FDA GUDID 06936415912284
- Registered EUDAMED B-06936415912284
- Recall initiated FDA recalls Z-2949-2024
- Recall initiated FDA recalls Z-2950-2024
- Recall initiated FDA recalls Z-2951-2024
- Recall initiated FDA recalls Z-2952-2024
- Recall initiated FDA recalls Z-0720-2026
- Recall initiated FDA recalls Z-0721-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Anesthesia System
- Generic name
- Anaesthesia unit, mobile
- Device name
- Anesthesia Machine
- Model / reference
- A8/A8C /A8 Pro/A80 /A80C/ A9/A9C /A9 Pro/A90 /A90C EUDAMEDA8/A9 EUDAMEDA9 GUDID
- Primary DI / UDI-DI
- 06936415912284
- Basic UDI-DI
- B-06936415912284
- 510(k) Number
- FDA product code
- BSZ
- EMDN code
- Z1203010101 Anaesthesia devices
- GMDN term
- Anaesthesia unit, mobile
- Device class
- Class IIb EUDAMEDClass II GUDID
- Regulation number
- 868.5160
- EU registration type
- Legacy device (under a directive)
- EU market of registration
- Germany
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Unsafe
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / reference | A8/A8C /A8 Pro/A80 /A80C/ A9/A9C /A9 Pro/A90 /A90C A8/A9 | A9 |
| Device class | Class IIb | Class II |
| Contains latex | No | Yes |
Registry differences
EUDAMED and FDA GUDID describe the same product (same UDI-DI) and state different values:
- Contains latex UDI-DI 06936415912284: EUDAMED states No, FDA GUDID states Yes.
Device class is stated on each registry's own scale: Class IIb in the EU (EUDAMED), Class II in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.
Manufacturer details
- EUDAMED SRN
- CN-MF-000014156
- Authorised representative
- Shanghai International Holding Corp. GmbH(Europe) DE-AR-000000001
Similar devices in this category
Other devices filed under ANAESTHESIA DEVICES.
- Accutron Inc. 26400-BWR by ACCUTRON
- Accutron Inc. 25615-FRU by ACCUTRON
- Prima 445 Anaesthetic Machine 5007370 by Penlon Limited
- Atlan A100 XL 8621200 by Drägerwerk AG & Co. KGaA
- Carestation 750c A1 1012-9755-000 by Datex-Ohmeda Inc.
- Fabius MRI 8607300 by Drägerwerk AG & Co. KGaA
- Prima 451 MRI Anaesthetic Machine 5011300 by Penlon Limited
- Carestation 650 A1 1012-9650-000 by Datex-Ohmeda Inc.
- Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia System by Maquet Critical Care AB
- Thornhill Research Inc. 124198 by Thornhill Research Inc
Other devices from Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
See all devices from Shenzhen Mindray Bio-Medical Electronics Co., Ltd. →