Fabius MRI 8607300
Ref 8607300Model 041498
Fabius MRI 8607300 is a device classified as Class IIb in the EU and Class II in the US made by Drägerwerk AG & Co. KGaA, registered in EUDAMED on 27 May 2026 (Basic UDI-DI B-04048675041498), cleared under FDA 510(k) K072884 on 30 Jul 2008 and listed in FDA GUDID since 24 Oct 2015 (UDI-DI 04048675041498), active as of 12 Nov 2026. 4 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
None ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDD EUDAMED | NumberB-04048675041498 04048675041498 | Date27 May 2026 | StatusActive |
| United States | GUDID listing GUDID | Number04048675041498 | Date24 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK072884 | Date30 Jul 2008 | StatusActive |
Official records
- EUDAMED public site B-04048675041498
- AccessGUDID record 04048675041498
- FDA 510(k) database K072884
Across registries
- EUDAMED Basic UDI-DI B-04048675041498 registered 27 May 2026 Active
- FDA 510(k) K072884 decided 30 Jul 2008 Active
- FDA GUDID UDI-DI 04048675041498 listed 24 Oct 2015 Listed
Recalls
-
- Terminated Z-0501-2022
26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.
-
- Terminated Z-2252-2018
The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
-
- Terminated Z-1481-2018
The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
-
- Terminated Z-2271-2015
the Fabius MRI or parts of the system were attracted by the magnetic field of the MRI. If the Fabius MRI is positioned too close to the MRI, its parts can be loosened as a result of the magnetic force of the MRI.
Timeline
- Cleared FDA 510(k) K072884
- Recall initiated FDA recalls Z-2271-2015
- Listed FDA GUDID 04048675041498
- Recall initiated FDA recalls Z-1481-2018
- Recall initiated FDA recalls Z-2252-2018
- Recall initiated FDA recalls Z-0501-2022
- Registered EUDAMED B-04048675041498
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Fabius MRI
- Generic name
- Anaesthesia workstation, general-purpose
- Model / reference
- 8607300, 041498
- Primary DI / UDI-DI
- 04048675041498
- Basic UDI-DI
- B-04048675041498
- FDA product code
- BSZ
- EMDN code
- Z1203010101 Anaesthesia devices
- GMDN term
- Anaesthesia workstation, general-purpose
- Device class
- Class IIb EUDAMEDClass II GUDID
- Regulation number
- 868.5160
- EU registration type
- Legacy device (under a directive)
- EU market of registration
- Germany
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Safe
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class IIb | Class II |
Similar devices in this category
Other devices filed under ANAESTHESIA DEVICES.
- Prima 440 Anaesthetic Machine 5007390 by Penlon Limited
- Anesthesia System A7 by Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
- Accutron Inc. 56500-UK by ACCUTRON
- Digital Ultra Flushmount Package G, INT 56600-INT by ACCUTRON
- Carestation 750 A1 1012-9750-000 by Datex-Ohmeda Inc.
- Anesthesia Machine A3 by Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
- TEC 6 PLUS VAP DES 220/240V DE SERVICED X1107-9604-000 by Datex-Ohmeda Inc.
- Thornhill Research Inc. 124198 by Thornhill Research Inc
- Prima 445 Anaesthetic Machine 5007370 by Penlon Limited
- Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia System by Maquet Critical Care AB