Apollo EZ
by Villa Sistemi Medicali S.p.A.
Part of a family of 34 devices: see the family page
Ref 9784NB0031
Apollo EZ (9784NB0031) is a device classified as Class IIb in the EU and Class II in the US made by Villa Sistemi Medicali S.P.A., registered in EUDAMED on 28 Jul 2025 (Basic UDI-DI 805773712FT2209C), cleared under FDA 510(k) K050190 on 25 Mar 2005 and listed in FDA GUDID since 24 Sep 2016 (UDI-DI 08057737120213), active as of 22 Oct 2026. 14 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
8 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number805773712FT2209C 08057737120213 | Date28 Jul 2025 | StatusActive |
| United States | GUDID listing GUDID | Number08057737120213 | Date24 Sep 2016 | StatusListed |
| United States | FDA 510(k) GUDID | NumberK050190 | Date25 Mar 2005 | StatusActive |
Official records
- EUDAMED public site 805773712FT2209C
- AccessGUDID record 08057737120213
- FDA 510(k) database K050190
Across registries
- EUDAMED Basic UDI-DI 805773712FT2209C registered 28 Jul 2025 Active
- FDA 510(k) K050190 decided 25 Mar 2005 Active
- FDA GUDID UDI-DI 08057737120213 listed 24 Sep 2016 Listed
Recalls
-
- Ongoing Z-0763-2020
- Ongoing Z-0765-2020
- Ongoing Z-0764-2020
- Ongoing Z-0766-2020
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
-
- Ongoing Z-1063-2017
- Ongoing Z-1064-2017
- Ongoing Z-1065-2017
- Ongoing Z-1066-2017
The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
-
- Terminated Z-1429-2017
- Terminated Z-1427-2017
- Terminated Z-1430-2017
- Terminated Z-1428-2017
Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.
-
- Terminated Z-1079-2015
The system does not provide the appropriate audible signal, permanent activation, and manual override, although the system is in high-level control functionality
-
- Terminated Z-2160-2013
It has been discovered that the Air Kerma rate information is inconsistent in Continuous Fluoro modes 30x30 (12"x12") and 15x15 (6"x6). This could result in a deviation of the displayed values of Air Kerma Rate and Cumulated Air Kerma from the actual values greater then +/-35% and then in a failure to comply with the requirements of 21 CFR 1020.32(k)(6
-
- Terminated Z-0702-2013
Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21
Timeline
- Listed FDA GUDID 08057737120213
- Recall initiated FDA recalls Z-1429-2017
- Recall initiated FDA recalls Z-1427-2017
- Recall initiated FDA recalls Z-1430-2017
- Recall initiated FDA recalls Z-1428-2017
- Recall initiated FDA recalls Z-0763-2020
- Recall initiated FDA recalls Z-0765-2020
- Recall initiated FDA recalls Z-0764-2020
- Recall initiated FDA recalls Z-0766-2020
- Registered EUDAMED 805773712FT2209C
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Apollo EZ EUDAMEDVilla Sistemi Medicali GUDID
- Generic name
- Stationary general-purpose fluoroscopic x-ray system, analogue
- Device name
- Apollo
- Model / reference
- 9784NB0031 EUDAMEDApollo EZ GUDID
- Primary DI / UDI-DI
- 08057737120213
- Basic UDI-DI
- 805773712FT2209C
- Direct marketing DI
- 08057737120213
- 510(k) Number
- FDA product code
- KXJ, IXL, KPW
- EMDN code
- Z11030701 Conventional remote control tables
- GMDN term
- Stationary general-purpose fluoroscopic x-ray system, analogue
- Device class
- Class IIb EUDAMEDClass II GUDID
- Regulation number
- 892.1980, 892.1670, 892.1610
- EU registration type
- Standard device
- EU market of registration
- Italy
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / reference | 9784NB0031 | Apollo EZ |
| Device class | Class IIb | Class II |
Manufacturer details
- EUDAMED SRN
- IT-MF-000016860
Similar devices in this category
Other devices filed under REMOTE CONTROL TABLES.
- X-RAY TV SYSTEM SONIALVISION G4 566-10000-13 by SHIMADZU CORPORATION
- uOmniscan 88002494 by Shanghai United Imaging Healthcare Co., Ltd.
- SONIALVISION G4 566-10000-42 by SHIMADZU CORPORATION
- NX 4.0 XCU8O by Agfa NV
- DR 800 XCTMA by Agfa NV
- Cardiovascular Angiography System by Allengers Medical Systems Limited