    • Terminated Z-0291-2013 FDA Class I initiated 13 Aug 2012
    • Terminated Z-0290-2013 FDA Class I initiated 13 Aug 2012

    Mindray has identified the potential for a system leak resulting an issue related to the carbon dioxide absorber canister gasket on A3/A5 Anesthesia Delivery Systems. A potential system leak resulting from the canister gasket will likely present itself during the Automatic Circuit Leak and Compliance Test performed at startup and the Manual Leak Test recommended to be performed before each case.