Endeavour Monitoring System
510(k) K102004
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK102004 | Date10 Sep 2010 | StatusActive |
Recalls
-
- Terminated Z-0389-2014
Mindray has identified an issue with V Series Monitor where the V Dock power pins connecting the V Series Monitor to the V Dock may lose spring tension, potentially causing the screen to temporarily go blank, lose communication with the central station monitor (when one is in use) or shut down. No patient related events have been reported to Mindray concerning the issue.
-
- Terminated Z-0528-2013
Mindray has identified an issue with the V Series Monitor where the monitors touch screen may stop responding to touch.
-
- Terminated Z-2279-2012
Mindray has identified two software anomalies contained in the V Series Monitoring System. There have been no reports of injuries associated with these anomalies. These anomalies affect V Series Systems distributed between September 31, 2010 and January 9, 2012.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MHX
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.1025
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent