Automated in Vitro Quantitative Assay for the Measurement of Allergen Specific Ige Antibodies
510(k) K122197
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK122197 | Date28 Aug 2013 | StatusActive |
Recalls
-
- Terminated Z-0947-2015
During an investigation of instrument logs it was determined that In specific circumstances involving multiple steps, a rack sequencing error may occur. This will result in a mismatch between the sample ID and the test result reported for all subsequent sample racks in that run.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DHB
- Device Class
- Class II
- Regulation
- 21 CFR Part 866.5750
- Advisory Committee
- Immunology
- FDA Decision
- SESE Substantially Equivalent