Skip to content

Automated in Vitro Quantitative Assay for the Measurement of Allergen Specific Ige Antibodies

by Phadia US, Inc.

510(k) K122197

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 10 Oct 2026. What this means

Regulatory status

as of registry sync 10 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK122197 Date28 Aug 2013ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-0947-2015 FDA Class II initiated 20 Nov 2014

    During an investigation of instrument logs it was determined that In specific circumstances involving multiple steps, a rack sequencing error may occur. This will result in a mismatch between the sample ID and the test result reported for all subsequent sample racks in that run.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DHB
Device Class
Class II
Regulation
21 CFR Part 866.5750
Advisory Committee
Immunology
FDA Decision
SESE Substantially Equivalent

Other devices from Phadia Us Inc.