Bard Mission Disposable Core Biopsy Instrument
by Bard Peripheral Vascular, Inc. 510(k)
Active
510(k) K171953
Regulatory status
Recall ongoing. 3 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK171953 | Date14 Sep 2017 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-0643-2024
The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures.
-
- Ongoing Z-1564-2023
Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue, which could lead to procedure delay, and insufficient sample acquisition requiring a repeat procedure.
-
- Ongoing Z-1260-2023
The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if used despite incompatibility.
Specification
- 510(k) Number
- FDA Product Code
- KNW
- Device Class
- Class II
- Regulation
- 21 CFR Part 876.1075
- Advisory Committee
- Gastroenterology, Urology
- FDA Decision
- SESE Substantially Equivalent