Bci Advisor Vital Signs Monitor (9200)
510(k) K011177
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK011177 | Date23 May 2001 | StatusActive |
Recalls
-
- Terminated Z-1326-2009
Recent improvements were made to the Advisor case plastic resulting in the inserts being deeply embedded into the plastic. This affects the ability of the monitor to be adequately secured on the wall mount bracket or roll stand bracket, resulting in the monitor falling off. There have been no adverse events or injuries reported.
-
- Terminated Z-0226-06
- Terminated Z-0229-06
- Terminated Z-0239-06
- Terminated Z-0221-06
- Terminated Z-0235-06
- Terminated Z-0223-06
- Terminated Z-0230-06
- Terminated Z-0225-06
- Terminated Z-0231-06
- Terminated Z-0227-06
- Terminated Z-0232-06
- Terminated Z-0228-06
- Terminated Z-0224-06
- Terminated Z-0238-06
- Terminated Z-0234-06
- Terminated Z-0237-06
- Terminated Z-0220-06
- Terminated Z-0233-06
- Terminated Z-0222-06
- Terminated Z-0240-06
- Terminated Z-0236-06
Potential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the primary lead is set to Lead II, III or any of the augmented leads, AVL, AVR or AVF on the Advisor monitor.
-
- Terminated Z-0828-05
- Terminated Z-0830-05
- Terminated Z-0824-05
- Terminated Z-0831-05
- Terminated Z-0827-05
- Terminated Z-0835-05
- Terminated Z-0834-05
- Terminated Z-0832-05
- Terminated Z-0826-05
- Terminated Z-0829-05
- Terminated Z-0833-05
- Terminated Z-0825-05
Altitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard temperature pressure). The % CO values displayed will be lower than the actual CO2 value.
-
- Terminated Z-0533-05
- Terminated Z-0534-05
- Terminated Z-0531-05
- Terminated Z-0535-05
- Terminated Z-0532-05
An issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DPS
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.2340
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent