Biomet 360 32-360137
Biomet 360 32-360137 is a Class II device made by Biomet Orthopedics, Inc., cleared under FDA 510(k) K093293 on 22 Oct 2010 and listed in FDA GUDID since 21 Sep 2016 (UDI-DI 00880304495920), active as of 23 Oct 2026. 5 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00880304495920 | Date21 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK093293 | Date22 Oct 2010 | StatusActive |
Official records
- AccessGUDID record 00880304495920
- FDA 510(k) database K093293
Across registries
- FDA 510(k) K093293 decided 22 Oct 2010 Active
- FDA GUDID UDI-DI 00880304495920 listed 21 Sep 2016 Listed
Recalls
-
- Terminated Z-0723-2021
Incorrect bolt -tibial bolts in package instead of femoral bolts may delay procedure
-
- Terminated Z-3249-2018
- Terminated Z-3250-2018
The implant only has one bolt in the sterile package but it should have had two.
-
- Terminated Z-0003-2017
The drill bit used to drill the hex pre-drilled hole in the screw head of the Tibial Augment Bolts included in the Vanguard 360 Revision Knee System was larger than allowable per the process specification. As a result, the hex geometry can still be visible, but there are radial cuts into the hex flat. Therefore, when the bolt is inserted into the augment, the screwdriver is unable to catch on the hex geometry, or the screwdriver strips the remaining hex geometry, and the bolt is unable to be fully tightened. The bolt stripping or the bolt not being fully tightened will be noticed by the surgeon intraoperatively when he or she is unable to screw the augment bolts into the augment. This possible event can result in an intraoperative delay resulting in increased time under anesthesia.
-
- Terminated Z-2242-2014
The Vanguard 360 Femoral Boss Reamers were manufactured oversized to the print. The instrument may not fit into the mating component.
Timeline
- Cleared FDA 510(k) K093293
- Recall initiated FDA recalls Z-2242-2014
- Recall initiated FDA recalls Z-0003-2017
- Listed FDA GUDID 00880304495920
- Recall initiated FDA recalls Z-3249-2018
- Recall initiated FDA recalls Z-3250-2018
- Recall initiated FDA recalls Z-0723-2021
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Biomet 360
- Generic name
- Knee tibia prosthesis trial, reusable
- Model / Ref / Catalog
- 32-360137
- Primary DI / UDI-DI
- 00880304495920
- 510(k) Number
- FDA product code
- MBV, JWH
- GMDN term
- Knee tibia prosthesis trial, reusable
- Device class
- Class II
- Regulation number
- 888.3560
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
Similar devices in this category
Other devices filed under Knee tibia prosthesis trial, reusable.
- Persona® 42-5892-007-30 by Zimmer, Inc.
- NexGen® 5881-40-14 00-5881-040-14 by Zimmer, Inc.
- U2 Total Knee System 2303-2402-RF by UNITED ORTHOPEDIC CORP.
- Unity Total Knee System 123.369.69 by CORIN LTD
- Unity Total Knee System 1236030 by CORIN LTD
- OMNIBotics NS-5016R by OMNIlife science, Inc.
- APEX Knee System KS-94611 by OMNIlife science, Inc.