Biomet 360 Knee System 185253
Biomet 360 Knee System 185253 is a Class II device made by Biomet Orthopedics, Inc., cleared under FDA 510(k) K093293 on 22 Oct 2010 and listed in FDA GUDID since 24 Oct 2015 (UDI-DI 00880304481886), active as of 23 Oct 2026. 5 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00880304481886 | Date24 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE | NumberK093293 | Date22 Oct 2010 | StatusActive |
Official records
- AccessGUDID record 00880304481886
- FDA 510(k) database K093293
Across registries
- FDA 510(k) K093293 decided 22 Oct 2010 Active
- FDA GUDID UDI-DI 00880304481886 listed 24 Oct 2015 Listed
Recalls
-
- Terminated Z-0723-2021
Incorrect bolt -tibial bolts in package instead of femoral bolts may delay procedure
-
- Terminated Z-3249-2018
- Terminated Z-3250-2018
The implant only has one bolt in the sterile package but it should have had two.
-
- Terminated Z-0003-2017
The drill bit used to drill the hex pre-drilled hole in the screw head of the Tibial Augment Bolts included in the Vanguard 360 Revision Knee System was larger than allowable per the process specification. As a result, the hex geometry can still be visible, but there are radial cuts into the hex flat. Therefore, when the bolt is inserted into the augment, the screwdriver is unable to catch on the hex geometry, or the screwdriver strips the remaining hex geometry, and the bolt is unable to be fully tightened. The bolt stripping or the bolt not being fully tightened will be noticed by the surgeon intraoperatively when he or she is unable to screw the augment bolts into the augment. This possible event can result in an intraoperative delay resulting in increased time under anesthesia.
-
- Terminated Z-2242-2014
The Vanguard 360 Femoral Boss Reamers were manufactured oversized to the print. The instrument may not fit into the mating component.
Timeline
- Cleared FDA 510(k) K093293
- Recall initiated FDA recalls Z-2242-2014
- Listed FDA GUDID 00880304481886
- Recall initiated FDA recalls Z-0003-2017
- Recall initiated FDA recalls Z-3249-2018
- Recall initiated FDA recalls Z-3250-2018
- Recall initiated FDA recalls Z-0723-2021
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Biomet 360 Knee System
- Generic name
- Knee arthroplasty wedge
- Model / Ref / Catalog
- 185253
- Primary DI / UDI-DI
- 00880304481886
- 510(k) Number
- FDA product code
- MBV, JWH
- GMDN term
- Knee arthroplasty wedge
- Device class
- Class II
- Regulation number
- 888.3560
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Conditional
Similar devices in this category
Other devices filed under Knee arthroplasty wedge.
- NexGen® 5988-01-29 00-5988-001-29 by Zimmer, Inc.
- Persona® Trabecular Metal® 42-5450-010-14 by Zimmer, Inc.
- APEX Revision Knee KP-43100 by OMNIlife science, Inc.
- Endo-Model Modular Knee Prosthesis System 15-2982/01 by WALDEMAR LINK GmbH & Co. KG
- GMK Hinge and Revision TiNbN coated 02.09.8TA605 by Medacta International
- GMK-Revision/Hinge Total Knee System 02.07.112FDA by Medacta International
- APEX Revision Knee KP-4515L by OMNIlife science, Inc.
- Unity Total Knee System 1123206 by CORIN LTD
- TrabecuLink Femoral Cones 15-3992/07 by WALDEMAR LINK GmbH & Co. KG