BLANKETROL III - Coolblue
by Cincinnati Sub-Zero Products, Inc.
510(k) K101589
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK101589 | Date20 Sep 2010 | StatusActive |
Recalls
-
- Ongoing Z-2970-2026
Overheating occurring on 115V CPU boards for the Blanketrol III Model 233, Innercool CoolBlue Model STx, and Blanketrol II Model 222S products may cause the device to shut down,
-
- Ongoing Z-2969-2026
Overheating occurring on 115V CPU boards for the Blanketrol III Model 233, Innercool CoolBlue Model STx, and Blanketrol II Model 222S products may cause the device to shut down.
-
- Terminated Z-0419-2020
- Terminated Z-0417-2020
Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added and clarifications were made regarding the use of the Automatic modes
-
- Terminated Z-2209-2013
During a label review it was discovered that both the Blanketrol III Operations (57201-N) & Operations & Technical (56201-Q) manuals had verbiage added back into the manual during ECN M111-4260 (released 12/29/2011) which was previously removed for clarification of the intended use and subsequent 501(K) approval.
-
- Terminated Z-1817-2011
The firm initiated this recall due to changes that have been made to the Blanketrol III Model 233 device's Operation and Operation/Technical Service Manuals, in order to stay in compliance with labeling regulations.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DWJ
- Models / Variants
- Coolblue
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.5900
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent