Brilliance Volume
by Philips Medical Systems (Cleveland), Inc.
510(k) K060937
Brilliance Volume is a medical device (FDA Class II; FDA product code JAK, System, X-Ray, Tomography, Computed) made by Philips Medical Systems (Cleveland), Inc., cleared under FDA 510(k) K060937 on 5 Jun 2006, active as of 22 Oct 2026. 58 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK060937 | Date5 Jun 2006 | StatusActive |
Official records
- FDA 510(k) database K060937
From the FDA record
- Received by the FDA
- 5 Apr 2006
- FDA decision
- 5 Jun 2006 (Substantially Equivalent)
- FDA review time
- 61 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-2599-2025
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
-
- Ongoing Z-1990-2021
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
-
- Terminated Z-1774-2021
Cooling Unit (CLU) expelled out of a CT gantry of an iCT system may result in injury
-
- Terminated Z-1921-2020
- Terminated Z-1922-2020
The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.
-
- Terminated Z-1409-2018
- Terminated Z-1408-2018
Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
-
- Terminated Z-0332-2018
- Terminated Z-0331-2018
There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry.
-
- Terminated Z-0403-2018
- Terminated Z-0402-2018
Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other functions.
-
- Terminated Z-1058-2017
During a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued.
-
- Terminated Z-1004-2017
Failure to correctly document the installation of four M12 Bolts into the system rotor.
-
- Terminated Z-0593-2017
- Terminated Z-0594-2017
The product label does not include the correct current rating.
-
- Terminated Z-2112-2016
Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack.
-
- Terminated Z-2381-2016
Software issues in versions v4.1 .3/4.1.5 in the Philips Brilliance iCT) iCT SP products that, could affect the performance of the equipment.
-
- Terminated Z-2382-2016
Software issues found in v4.1 .3/4.1.5 in the Philips Brilliance iCT/ iCT SP products that could affect the performance of the equipment.
-
- Terminated Z-1506-2016
The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90 or 180 relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.
-
- Terminated Z-1352-2016
- Terminated Z-1353-2016
The firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.
-
- Terminated Z-0545-2016
- Terminated Z-0546-2016
Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not access the basic axial perfusion feature after upgrading to software version 4.1.2.
-
- Terminated Z-1713-2016
- Terminated Z-1715-2016
- Terminated Z-1714-2016
Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm
-
- Terminated Z-1695-2015
A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.
-
- Terminated Z-1656-2015
The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.
-
- Terminated Z-1767-2015
- Terminated Z-1768-2015
- Terminated Z-1769-2015
The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
-
- Terminated Z-2256-2015
- Terminated Z-2255-2015
5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value incorrectly remains constant for some heart rate change conditions. DoseRight incorrectly enabled when the Exam Card specified disabling DoseRight. Tube heat predict
-
- Terminated Z-1359-2015
Philips discovered that a software defect exists in marketed product wherein the sign indication of the longitudinal position of some types of scan is inverted.
-
- Terminated Z-0979-2015
Improper assembly of the vertical motor/brake system may lead to uncommanded vertical motion.
-
- Terminated Z-0236-2015
It was discovered that a software defect may result in the scanner not terminating the CT scan at the intended location.
-
- Terminated Z-2237-2014
The following software issues have been identified in the affected products. Problem 1: When reopening a bookmark generated from processing a MUGA (Multi-Gated Acquisition) scan within the NM Cardiac Application, the ejection fraction (EF) displayed should be the same as the ejection fraction (EF) originally displayed when the bookmark was first created. In some instances, the ejection fraction (
-
- Terminated Z-1722-2014
A customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.
-
- Terminated Z-1461-2014
Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.
-
- Terminated Z-1678-2014
During SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not applied correctly in SPECT reconstructions using Astonish in AutoSPECT Pro on Intellispace Portal.
-
- Terminated Z-0832-2014
Patient images exhibited ring artifacts.
-
- Terminated Z-0335-2014
If the couch, with the footrest extension attached, is positioned where the table travel is within the bore of the gantry, and the user begins to lower the couch, the footrest extension can contact the gantry cover or come in contact with the user due to the length of the accessory.
-
- Terminated Z-0831-2014
An artifact that resembles thrombus may appear on the image.
-
- Terminated Z-0945-2013
Philips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.
-
- Terminated Z-2035-2012
Philips Healthcare was notified that several of the sixteen bolts securing the brackets which support the carbon table top on various models of their Brilliant CT scanner have been found to be defective and the bolt heads may break off.
-
- Terminated Z-1988-2012
Philips is issuing a software update to the Intellispace Portal systems to address a number of non-conformances. The most frequent and severe non-conformance identified involves the portal setting of "PAC Integration". This setting has a feature that allows the user to launch images of multiple patient studies into an application, but does not display a warning message informing the user that diff
-
- Terminated Z-1987-2012
The Intellispace Portal software and Extended Brillance Workspace software may have significantly different results in the colors and size of the affected area when the customer uses the "Time Sensitive" and "Time Insensitive" methods.
-
- Terminated Z-2000-2012
In the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion measurements.
-
- Terminated Z-1688-2012
On 3/5/2012, Philips Healthcare was notified of one incident involving a Brilliance iCT scanner that resulted in damage to the gantry and slight injury to the user when the heat exchanger detached from the cooling unit due to some missing welds.
-
- Terminated Z-1791-2012
Philips is issuing a software update (version 3.2.4) to the Brilliance iCT and iCT SP systems currently running software version 3.2.0, 3.2.1 or 3.2.3, addressing a number of non-conformances. A review was conducted of all nonconformances that are fixed by the software upgrade. None of these nonconformances result in unacceptable hazards. The nonconformance documented below had the highest sc
-
- Terminated Z-1693-2012
- Terminated Z-1691-2012
- Terminated Z-1692-2012
Component defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems were not manufactured to specification and may crack.
-
- Terminated Z-1337-2012
A number of non-conformances were logged for IntelliSpace Portal running software version 4.0 during in-house random verification testing. When loading a study to Review, after failing to load it through Add to Running Application the Analysis application is not being closed and different (1st loaded study) patient appears in analysis (CCA).
-
- Terminated Z-0845-2012
The Common Image Reconstruction System (CIRS) on the affected Brilliance Systems encounters an error and requires reboot of the system. The unit failed to initialize axial scan due to timeout that occurred waiting for "READY" state
-
- Terminated Z-3208-2011
- Terminated Z-3209-2011
Software errors. Philips Medical's Field Service Engineers are installing software upgrades to certain of their Brilliance iCT and Brilliance ICT SP CT X-ray system units in order to correct non-conformances evidenced in these systems due to software errors.
-
- Terminated Z-0791-2014
The DoseRight feature suggest a mAs based on the measured patient size, a reference size and a reference mAs. When scanning large children, the suggested mAs may be higher than clinicians would expect.
-
- Terminated Z-2205-2011
A program software error in the algorithm produces image artifacts that could possibly lead to a misdiagnosis of patients who have small metal implants.
-
- Terminated Z-1669-2011
Philips Medical Systems has decided to recall the patient support vertical brake hub in all its Brilliance systems due to the collapse in the vertical direction. The short key connecting the vertical brake hub to the motor shaft can disengage due to the lack of thread locker on set screws.
-
- Terminated Z-2460-2010
The gantry covers of the iCT have a latch at the base of the system which holds the cover in the closed position. During the course of maintenance the CT Field Service Engineer (FSE) will activate this latching mechanism which in turn has a potential to penetrate the insulation on cabling which carries 277 VAC. If the latch does come into contact with the cabling there is a potential that the FS
Timeline
- Recall initiated FDA recalls Z-0402-2018
- Recall initiated FDA recalls Z-0332-2018
- Recall initiated FDA recalls Z-0331-2018
- Recall initiated FDA recalls Z-1409-2018
- Recall initiated FDA recalls Z-1408-2018
- Recall initiated FDA recalls Z-1921-2020
- Recall initiated FDA recalls Z-1922-2020
- Recall initiated FDA recalls Z-1774-2021
- Recall initiated FDA recalls Z-1990-2021
- Recall initiated FDA recalls Z-2599-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- JAK
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1750
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code JAK (System, X-Ray, Tomography, Computed).
- Cerebral Blood Flow Analysis System (CBP), Model CSCP-001A by Toshiba America Medical Systems, In.C
- Dental Digital PANORAMIC/3D Imaging System by E-Woo Technology
- NMCT 870CZT by GE Medical Systems Israel, Functional Imaging
- CT COLONOGRAPHY/NAVIGATOR2 by GE MEDICAL SYSTEMS, INC.
- CT Perfusion 2 by General Electric Co.
- DeltaScan 2000 Series by Ohio Nuclear, Inc.
- syngo.CT Dual Energy by SIEMENS MEDICAL SOLUTIONS USA, INC.
- Aquilion ONE (TSX-305A/6) V8.9 with AiCE by Canon Medical Systems Corporation
- MaxFOV 2 by GE Medical Systems, LLC
- Sytec Dentascan by General Electric Co.
Other devices from Philips Medical Systems (Cleveland), Inc.
See all devices from Philips Medical Systems (Cleveland), Inc. →