Skip to content

Brilliance Volume

by Philips Medical Systems (Cleveland), Inc.

510(k) K060937

Brilliance Volume is a medical device (FDA Class II; FDA product code JAK, System, X-Ray, Tomography, Computed) made by Philips Medical Systems (Cleveland), Inc., cleared under FDA 510(k) K060937 on 5 Jun 2006, active as of 22 Oct 2026. 58 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK060937 Date5 Jun 2006ClearedStatusActive

Official records

From the FDA record

Received by the FDA
5 Apr 2006
FDA decision
5 Jun 2006 (Substantially Equivalent)
FDA review time
61 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K060937 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

61 recalls · 2 ongoing · FDA

    • Ongoing Z-2599-2025 FDA Class II initiated 13 Aug 2025

    The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

    • Ongoing Z-1990-2021 FDA Class II initiated 1 May 2021

    Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

    • Terminated Z-1774-2021 FDA Class II initiated 13 Apr 2021

    Cooling Unit (CLU) expelled out of a CT gantry of an iCT system may result in injury

  • 2 recalls, same reason

    • Terminated Z-1921-2020 FDA Class II initiated 6 Apr 2020
    • Terminated Z-1922-2020 FDA Class II initiated 6 Apr 2020

    The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.

  • 2 recalls, same reason

    • Terminated Z-1409-2018 FDA Class II initiated 8 Feb 2018
    • Terminated Z-1408-2018 FDA Class II initiated 8 Feb 2018

    Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.

  • 2 recalls, same reason

    • Terminated Z-0332-2018 FDA Class II initiated 8 Nov 2017
    • Terminated Z-0331-2018 FDA Class II initiated 8 Nov 2017

    There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry.

  • 2 recalls, same reason

    • Terminated Z-0403-2018 FDA Class II initiated 11 Jul 2017
    • Terminated Z-0402-2018 FDA Class II initiated 11 Jul 2017

    Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other functions.

    • Terminated Z-1058-2017 FDA Class II initiated 23 Dec 2016

    During a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued.

    • Terminated Z-1004-2017 FDA Class II initiated 25 Nov 2016

    Failure to correctly document the installation of four M12 Bolts into the system rotor.

  • 2 recalls, same reason

    • Terminated Z-0593-2017 FDA Class III initiated 13 Oct 2016
    • Terminated Z-0594-2017 FDA Class III initiated 13 Oct 2016

    The product label does not include the correct current rating.

    • Terminated Z-2112-2016 FDA Class II initiated 21 Jun 2016

    Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack.

    • Terminated Z-2381-2016 FDA Class II initiated 29 Apr 2016

    Software issues in versions v4.1 .3/4.1.5 in the Philips Brilliance iCT) iCT SP products that, could affect the performance of the equipment.

    • Terminated Z-2382-2016 FDA Class II initiated 29 Apr 2016

    Software issues found in v4.1 .3/4.1.5 in the Philips Brilliance iCT/ iCT SP products that could affect the performance of the equipment.

    • Terminated Z-1506-2016 FDA Class II initiated 4 Mar 2016

    The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90 or 180 relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.

  • 2 recalls, same reason

    • Terminated Z-1352-2016 FDA Class II initiated 5 Feb 2016
    • Terminated Z-1353-2016 FDA Class II initiated 5 Feb 2016

    The firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.

  • 2 recalls, same reason

    • Terminated Z-0545-2016 FDA Class II initiated 29 Oct 2015
    • Terminated Z-0546-2016 FDA Class II initiated 29 Oct 2015

    Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not access the basic axial perfusion feature after upgrading to software version 4.1.2.

  • 3 recalls, same reason

    • Terminated Z-1713-2016 FDA Class II initiated 6 Sep 2015
    • Terminated Z-1715-2016 FDA Class II initiated 6 Sep 2015
    • Terminated Z-1714-2016 FDA Class II initiated 6 Sep 2015

    Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

    • Terminated Z-1695-2015 FDA Class II initiated 13 Apr 2015

    A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.

    • Terminated Z-1656-2015 FDA Class II initiated 16 Mar 2015

    The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.

  • 3 recalls, same reason

    • Terminated Z-1767-2015 FDA Class II initiated 6 Mar 2015
    • Terminated Z-1768-2015 FDA Class II initiated 6 Mar 2015
    • Terminated Z-1769-2015 FDA Class II initiated 6 Mar 2015

    The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.

  • 2 recalls, same reason

    • Terminated Z-2256-2015 FDA Class II initiated 5 Feb 2015
    • Terminated Z-2255-2015 FDA Class II initiated 5 Feb 2015

    5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value incorrectly remains constant for some heart rate change conditions. DoseRight incorrectly enabled when the Exam Card specified disabling DoseRight. Tube heat predict

    • Terminated Z-1359-2015 FDA Class II initiated 5 Feb 2015

    Philips discovered that a software defect exists in marketed product wherein the sign indication of the longitudinal position of some types of scan is inverted.

    • Terminated Z-0979-2015 FDA Class II initiated 8 Dec 2014

    Improper assembly of the vertical motor/brake system may lead to uncommanded vertical motion.

    • Terminated Z-0236-2015 FDA Class II initiated 31 Jul 2014

    It was discovered that a software defect may result in the scanner not terminating the CT scan at the intended location.

    • Terminated Z-2237-2014 FDA Class II initiated 27 Jul 2014

    The following software issues have been identified in the affected products. Problem 1: When reopening a bookmark generated from processing a MUGA (Multi-Gated Acquisition) scan within the NM Cardiac Application, the ejection fraction (EF) displayed should be the same as the ejection fraction (EF) originally displayed when the bookmark was first created. In some instances, the ejection fraction (

    • Terminated Z-1722-2014 FDA Class II initiated 1 Apr 2014

    A customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.

    • Terminated Z-1461-2014 FDA Class II initiated 27 Mar 2014

    Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.

    • Terminated Z-1678-2014 FDA Class II initiated 11 Mar 2014 on shared 510(k) K060937

    During SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not applied correctly in SPECT reconstructions using Astonish in AutoSPECT Pro on Intellispace Portal.

    • Terminated Z-0832-2014 FDA Class II initiated 9 Dec 2013

    Patient images exhibited ring artifacts.

    • Terminated Z-0335-2014 FDA Class II initiated 11 Oct 2013

    If the couch, with the footrest extension attached, is positioned where the table travel is within the bore of the gantry, and the user begins to lower the couch, the footrest extension can contact the gantry cover or come in contact with the user due to the length of the accessory.

    • Terminated Z-0831-2014 FDA Class II initiated 16 Jul 2013

    An artifact that resembles thrombus may appear on the image.

    • Terminated Z-0945-2013 FDA Class II initiated 20 Dec 2012

    Philips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.

    • Terminated Z-2035-2012 FDA Class II initiated 2 Jul 2012

    Philips Healthcare was notified that several of the sixteen bolts securing the brackets which support the carbon table top on various models of their Brilliant CT scanner have been found to be defective and the bolt heads may break off.

    • Terminated Z-1988-2012 FDA Class II initiated 23 May 2012

    Philips is issuing a software update to the Intellispace Portal systems to address a number of non-conformances. The most frequent and severe non-conformance identified involves the portal setting of "PAC Integration". This setting has a feature that allows the user to launch images of multiple patient studies into an application, but does not display a warning message informing the user that diff

    • Terminated Z-1987-2012 FDA Class II initiated 23 May 2012

    The Intellispace Portal software and Extended Brillance Workspace software may have significantly different results in the colors and size of the affected area when the customer uses the "Time Sensitive" and "Time Insensitive" methods.

    • Terminated Z-2000-2012 FDA Class II initiated 22 May 2012

    In the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion measurements.

    • Terminated Z-1688-2012 FDA Not yet classified initiated 7 May 2012

    On 3/5/2012, Philips Healthcare was notified of one incident involving a Brilliance iCT scanner that resulted in damage to the gantry and slight injury to the user when the heat exchanger detached from the cooling unit due to some missing welds.

    • Terminated Z-1791-2012 FDA Class II initiated 7 May 2012

    Philips is issuing a software update (version 3.2.4) to the Brilliance iCT and iCT SP systems currently running software version 3.2.0, 3.2.1 or 3.2.3, addressing a number of non-conformances. A review was conducted of all nonconformances that are fixed by the software upgrade. None of these nonconformances result in unacceptable hazards. The nonconformance documented below had the highest sc

  • 3 recalls, same reason

    • Terminated Z-1693-2012 FDA Not yet classified initiated 26 Mar 2012
    • Terminated Z-1691-2012 FDA Not yet classified initiated 26 Mar 2012
    • Terminated Z-1692-2012 FDA Not yet classified initiated 26 Mar 2012

    Component defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems were not manufactured to specification and may crack.

    • Terminated Z-1337-2012 FDA Not yet classified initiated 22 Feb 2012 on shared 510(k) K060937

    A number of non-conformances were logged for IntelliSpace Portal running software version 4.0 during in-house random verification testing. When loading a study to Review, after failing to load it through Add to Running Application the Analysis application is not being closed and different (1st loaded study) patient appears in analysis (CCA).

    • Terminated Z-0845-2012 FDA Unclassified (pre-2012) initiated 7 Oct 2011

    The Common Image Reconstruction System (CIRS) on the affected Brilliance Systems encounters an error and requires reboot of the system. The unit failed to initialize axial scan due to timeout that occurred waiting for "READY" state

  • 2 recalls, same reason

    • Terminated Z-3208-2011 FDA Unclassified (pre-2012) initiated 15 Jun 2011
    • Terminated Z-3209-2011 FDA Unclassified (pre-2012) initiated 15 Jun 2011

    Software errors. Philips Medical's Field Service Engineers are installing software upgrades to certain of their Brilliance iCT and Brilliance ICT SP CT X-ray system units in order to correct non-conformances evidenced in these systems due to software errors.

    • Terminated Z-0791-2014 FDA Class II initiated 7 Mar 2011

    The DoseRight feature suggest a mAs based on the measured patient size, a reference size and a reference mAs. When scanning large children, the suggested mAs may be higher than clinicians would expect.

    • Terminated Z-2205-2011 FDA Unclassified (pre-2012) initiated 7 Mar 2011

    A program software error in the algorithm produces image artifacts that could possibly lead to a misdiagnosis of patients who have small metal implants.

    • Terminated Z-1669-2011 FDA Unclassified (pre-2012) initiated 10 Dec 2010 on shared 510(k) K060937

    Philips Medical Systems has decided to recall the patient support vertical brake hub in all its Brilliance systems due to the collapse in the vertical direction. The short key connecting the vertical brake hub to the motor shaft can disengage due to the lack of thread locker on set screws.

    • Terminated Z-2460-2010 FDA Unclassified (pre-2012) initiated 14 Apr 2010

    The gantry covers of the iCT have a latch at the base of the system which holds the cover in the closed position. During the course of maintenance the CT Field Service Engineer (FSE) will activate this latching mechanism which in turn has a potential to penetrate the insulation on cabling which carries 277 VAC. If the latch does come into contact with the cabling there is a potential that the FS

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

62 dated events

  1. Recall initiated FDA recalls Z-0402-2018
  2. Recall initiated FDA recalls Z-0332-2018
  3. Recall initiated FDA recalls Z-0331-2018
  4. Recall initiated FDA recalls Z-1409-2018
  5. Recall initiated FDA recalls Z-1408-2018
  6. Recall initiated FDA recalls Z-1921-2020
  7. Recall initiated FDA recalls Z-1922-2020
  8. Recall initiated FDA recalls Z-1774-2021
  9. Recall initiated FDA recalls Z-1990-2021
  10. Recall initiated FDA recalls Z-2599-2025

52 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
JAK
Device class
Class II
Regulation
21 CFR Part 892.1750
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code JAK (System, X-Ray, Tomography, Computed).

Other devices from Philips Medical Systems (Cleveland), Inc.

See all devices from Philips Medical Systems (Cleveland), Inc. →