Skip to content

Brivo Nm 615

by GE Healthcare

510(k) K113350

Brivo Nm 615 is a medical device (FDA Class II; FDA product code KPS, System, Tomography, Computed, Emission) made by GE Healthcare, cleared under FDA 510(k) K113350 on 5 Dec 2011, active as of 22 Oct 2026. 4 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK113350 Date5 Dec 2011ClearedStatusActive

Official records

From the FDA record

Received by the FDA
14 Nov 2011
FDA decision
5 Dec 2011 (Substantially Equivalent)
FDA review time
21 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K113350 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

4 recalls · 2 ongoing · FDA

    • Ongoing Z-0889-2025 FDA Class II initiated 18 Dec 2024

    GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

    • Ongoing Z-1055-2023 FDA Class I initiated 19 Dec 2022

    A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.

    • Terminated Z-0367-2014 FDA Class II initiated 25 Oct 2013

    GE Healthcare has recently become aware of a potential safety issue related to a SPECT scan procedure used for pulmonary imaging. Contact between the Detector (GE Healthcare Discovery NM/CT 670, and/or Optima NM/CT 640, and/or Discovery NM630, and/or Brivo NM 615 all configurations) and the Patient's elbow may occur during SPECT protocols using 450 angular steps in scenarios when a Patients a

    • Terminated Z-1790-2013 FDA Class I initiated 13 Jun 2013

    GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re

Timeline

5 dated events

  1. Cleared FDA 510(k) K113350
  2. Recall initiated FDA recalls Z-1790-2013
  3. Recall initiated FDA recalls Z-0367-2014
  4. Recall initiated FDA recalls Z-1055-2023
  5. Recall initiated FDA recalls Z-0889-2025

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
KPS
Device class
Class II
Regulation
21 CFR Part 892.1200
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Devices cleared under K113350

2 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K113350
UDI-DISize and differences
Brivo 615 5376496-15
GE Hualun Medical Systems Co., Ltd
5376496-1500840682117869 —On the market
Brivo NM 615
GE Hualun Medical Systems Co., Ltd
NM 61500840682139267 —On the market

Similar devices

Other devices filed under FDA product code KPS (System, Tomography, Computed, Emission).

Other devices from GE Healthcare

See all devices from GE Healthcare →