Brivo Nm 615
510(k) K113350
Brivo Nm 615 is a medical device (FDA Class II; FDA product code KPS, System, Tomography, Computed, Emission) made by GE Healthcare, cleared under FDA 510(k) K113350 on 5 Dec 2011, active as of 22 Oct 2026. 4 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK113350 | Date5 Dec 2011 | StatusActive |
Official records
- FDA 510(k) database K113350
From the FDA record
- Received by the FDA
- 14 Nov 2011
- FDA decision
- 5 Dec 2011 (Substantially Equivalent)
- FDA review time
- 21 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-0889-2025
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
-
- Ongoing Z-1055-2023
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
-
- Terminated Z-0367-2014
GE Healthcare has recently become aware of a potential safety issue related to a SPECT scan procedure used for pulmonary imaging. Contact between the Detector (GE Healthcare Discovery NM/CT 670, and/or Optima NM/CT 640, and/or Discovery NM630, and/or Brivo NM 615 all configurations) and the Patient's elbow may occur during SPECT protocols using 450 angular steps in scenarios when a Patients a
-
- Terminated Z-1790-2013
GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
Timeline
- Cleared FDA 510(k) K113350
- Recall initiated FDA recalls Z-1790-2013
- Recall initiated FDA recalls Z-0367-2014
- Recall initiated FDA recalls Z-1055-2023
- Recall initiated FDA recalls Z-0889-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- KPS
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1200
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K113350
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| Brivo 615 5376496-15 | 5376496-15 | 00840682117869 | — | On the market |
| Brivo NM 615 | NM 615 | 00840682139267 | — | On the market |
Similar devices
Other devices filed under FDA product code KPS (System, Tomography, Computed, Emission).
- MODIFICATION TO SCINTRON IV by Miegmbh
- MG ATC: ATTENUATION CORRECTION SYSTEM FOR DUAL-HEAD VARIABLE-ANGLE GAMMA CAMERA by Ge Medical Systems F.I. Haifa
- M-LINK by Medimage, Inc.
- QUEST PET IMAGING SYSTEM by Ugm Medical Systems, Inc.
- PHAROS by Brightonix Imaging
- SCATTER CORRECTION VOLUME SOFTWARE MODEL NSSC-020A by Toshiba Medical Systems
- Symbia VA10A Family by SIEMENS MEDICAL SOLUTIONS USA, INC.
- NEUROPET by Photodetection Systems, Inc.
- BCC GAMMA CAMERA by Is2 Research, Inc.
- ASPECT GAMMA CAMERA by Digital Scintigraphics, Inc.