Discovery Nm 630
510(k) K111445
Discovery Nm 630 is a medical device (FDA Class II; FDA product code KPS, System, Tomography, Computed, Emission) made by GE Healthcare, cleared under FDA 510(k) K111445 on 3 Jun 2011, active as of 22 Oct 2026. 9 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK111445 | Date3 Jun 2011 | StatusActive |
Official records
- FDA 510(k) database K111445
From the FDA record
- Received by the FDA
- 24 May 2011
- FDA decision
- 3 Jun 2011 (Substantially Equivalent)
- FDA review time
- 10 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-0890-2025
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
-
- Ongoing Z-1054-2023
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
-
- Terminated Z-0856-2021
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
-
- Terminated Z-2994-2020
- Terminated Z-2995-2020
Rotor bearing screws were found loose on detector.
-
- Terminated Z-1257-2019
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
-
- Terminated Z-0367-2014
GE Healthcare has recently become aware of a potential safety issue related to a SPECT scan procedure used for pulmonary imaging. Contact between the Detector (GE Healthcare Discovery NM/CT 670, and/or Optima NM/CT 640, and/or Discovery NM630, and/or Brivo NM 615 all configurations) and the Patient's elbow may occur during SPECT protocols using 450 angular steps in scenarios when a Patients a
-
- Terminated Z-1791-2013
GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
-
- Terminated Z-0121-2013
GE Healthcare has recently become aware of a potential safety issue associated with the table pallet of your Discovery NM/CT 670 and/or Discovery NM630 system which may impact operation safety. At scan setup, during gantry rotation towards scanning position, a potential contact between the table-pallet-edge and the detector was identified. Such occurrence may damage the pallet front-edge and
-
- Terminated Z-1111-2012
GE Healthcare has recently become aware of a potential safety issue associated with the collimator cart and lead cover of your Discovery NM/CT 670 and/or Discovery NM 630 system that may impact operator safety. 1. The cart locking pin release and cart disengagement could result in the collimator falling down and could lead to operator or service personnel injury. 2. Another issue is that the
Timeline
- Recall initiated FDA recalls Z-1111-2012
- Recall initiated FDA recalls Z-0121-2013
- Recall initiated FDA recalls Z-1791-2013
- Recall initiated FDA recalls Z-0367-2014
- Recall initiated FDA recalls Z-1257-2019
- Recall initiated FDA recalls Z-2994-2020
- Recall initiated FDA recalls Z-2995-2020
- Recall initiated FDA recalls Z-0856-2021
- Recall initiated FDA recalls Z-1054-2023
- Recall initiated FDA recalls Z-0890-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- KPS
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1200
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K111445
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| D630 Upgrade H2401JF | H2401JF | 00840682123761 | — | On the market |
| D630 Upgrade H2401XF | H2401XF / H2401XF | 00840682145503 | — | On the market |
| Discovery 630 H3100JY | H3100JY | 00840682115292 | — | On the market |
| Discovery NM 630 | NM 630 | 00840682139274 | — | On the market |
| NM 830 | 5376204-33 / 830 | 00195278788252 | — | On the market |
| NM 830 | 830 | 00840682140829 | — | On the market |
Similar devices
Other devices filed under FDA product code KPS (System, Tomography, Computed, Emission).
- BEACON, MODEL 211060 by Philips Medical Systems (Cleveland), Inc.
- CORT-EX by General Electric Co.
- MODIFICATION TO WBR, MODEL HR by U.C.G. Technologies, Ltd.
- C-QUEST by Medx, Inc.
- Symbia 6.5 by SIEMENS MEDICAL SOLUTIONS USA, INC.
- HITACHI SPECTRADIGITAL SERIES GAMMA CAMERA SYSTEMS by Hitachi Medical Corp. of America
- Cartesion Prime (PCD-1000A/3) V10.15 by Canon Medical Systems Corporation
- ECAT PRT by Cti Pet Systems, Inc.
- OPTIMA NUCLEAR TOMOGRAPHY STYSTEM by General Electric Co.
- MOBILE COMPACT DIGITAL CARDIAC CAMERA by Is2 Research, Inc.