Buretrol Solution Sets

by Baxter Healthcare Corp

510(k) K984381

Buretrol Solution Sets is a medical device (FDA Class II; FDA product code FPA, Set, Administration, Intravascular) made by Baxter Healthcare Corp, cleared under FDA 510(k) K984381 on 19 Feb 1999, active as of 22 Oct 2026. 13 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

16 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK984381 Date19 Feb 1999ClearedStatusActive

Official records

From the FDA record

Received by the FDA
8 Dec 1998
FDA decision
19 Feb 1999 (Substantially Equivalent)
FDA review time
73 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K984381 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

16 recalls · FDA

  • 4 recalls, same reason

    • Terminated Z-0012-2013 FDA Class I initiated 7 Sep 2012
    • Terminated Z-0013-2013 FDA Class I initiated 7 Sep 2012
    • Terminated Z-0014-2013 FDA Class I initiated 7 Sep 2012
    • Terminated Z-0015-2013 FDA Class I initiated 7 Sep 2012

    Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.

    • Terminated Z-0070-2011 FDA Unclassified (pre-2012) initiated 3 Sep 2010

    There are leaks in the Drip Chamber Check Valve Subassembly of the Solution Set. A leak in a sterile pathway could allow microbial entry, which could lead to a blood stream infection.

    • Terminated Z-0487-2010 FDA Unclassified (pre-2012) initiated 3 Nov 2009

    The tubing may separate from the distal Clearlink Y-site of the set, which could result in patient injury if the separation occurs during patient use.

  • 3 recalls, same reason

    • Terminated Z-0654-06 FDA Unclassified (pre-2012) initiated 15 Dec 2005 on shared 510(k) K984381
    • Terminated Z-0655-06 FDA Unclassified (pre-2012) initiated 15 Dec 2005 on shared 510(k) K984381
    • Terminated Z-0656-06 FDA Unclassified (pre-2012) initiated 15 Dec 2005 on shared 510(k) K984381

    The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.

    • Terminated Z-0199-06 FDA Unclassified (pre-2012) initiated 17 Mar 2005

    The Buretrol solution sets are currently labeled as acceptable for use with the Colleague Volumertric Pump, but there is a potential for a non-detection of an upstream occlusion under certain conditions.

    • Terminated Z-0845-05 FDA Unclassified (pre-2012) initiated 28 Feb 2005

    Possible occlusion of the Clearlink injection site valve due to inadequate silicone lubrication of the slit in the upper seal of the valve.

    • Terminated Z-0193-05 FDA Unclassified (pre-2012) initiated 28 Apr 2004

    Inadequate directions for use may have contributed to an increase in infection rates at a small number of customers beginning use of the Clearlink needleless device.

  • 4 recalls, same reason

    • Terminated Z-0504-03 FDA Unclassified (pre-2012) initiated 17 Oct 2002
    • Terminated Z-0508-03 FDA Unclassified (pre-2012) initiated 17 Oct 2002
    • Terminated Z-0510-03 FDA Unclassified (pre-2012) initiated 17 Oct 2002
    • Terminated Z-0523-03 FDA Unclassified (pre-2012) initiated 17 Oct 2002

    The I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

17 dated events

  1. Recall initiated FDA recalls Z-0199-06
  2. Recall initiated FDA recalls Z-0654-06
  3. Recall initiated FDA recalls Z-0655-06
  4. Recall initiated FDA recalls Z-0656-06
  5. Recall initiated FDA recalls Z-0487-2010
  6. Recall initiated FDA recalls Z-0070-2011
  7. Recall initiated FDA recalls Z-0012-2013
  8. Recall initiated FDA recalls Z-0013-2013
  9. Recall initiated FDA recalls Z-0014-2013
  10. Recall initiated FDA recalls Z-0015-2013

7 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
FPA
Device class
Class II
Regulation
21 CFR Part 880.5440
Advisory committee
General Hospital
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code FPA (Set, Administration, Intravascular).

Other devices from Baxter Healthcare Corp

See all devices from Baxter Healthcare Corp →