Skip to content

C253A

by FUJIFILM Corporation

Ref 04573596214510

C253A (04573596214510) is an extracorporeal ultrasound imaging transducer (EU MDR Class IIa; FDA Class II) made by Fujifilm Corporation, registered in EUDAMED on 13 Feb 2026 (Basic UDI-DI 457359621C253A46), cleared under FDA 510(k) K220295 on 29 Apr 2022 (3 earlier submissions) and listed in FDA GUDID since 22 Jan 2024 (UDI-DI 04573596214510), active as of 22 Oct 2026. 2 FDA recalls are linked to it.

Active European Union and United States. Also listed in GUDID (US).

CE MDR 510(k)

Valid until

No date given

CE: no date in EUDAMED. FDA: not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

Trust

Unverified

Source EUDAMED, GUDID, synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number457359621C253A46 Basic UDI-DI 04573596214510UDI-DIDate13 Feb 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number04573596214510 UDI-DI Date22 Jan 2024ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE GUDID from GUDID Substantially Equivalent · Abbreviated NumberK220295 Date29 Apr 2022DecidedStatusActive

Official records

From the FDA record

Device description
Probe
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 29 May 2026, as the labeller entered it.

Across registries

3 registries list this device

  • EUDAMED Basic UDI-DI 457359621C253A46 registered 13 Feb 2026 Active
  • FDA 510(k) K220295 decided 29 Apr 2022 Active· +3 more in the status table
  • FDA GUDID UDI-DI 04573596214510 listed 22 Jan 2024 Listed

Recalls

2 recalls · 2 ongoing · FDA

    • Ongoing Z-0189-2022 FDA Class II initiated 14 Oct 2021

    Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.

    • Ongoing Z-0192-2022 FDA Class II initiated 14 Oct 2021

    When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after performing an echocardiogram. The incorrect value is displayed. In addition, the error appears in the instruction manual

Timeline

8 dated events

  1. Cleared FDA 510(k) K181376
  2. Cleared FDA 510(k) K191233
  3. Cleared FDA 510(k) K202422
  4. Recall initiated FDA recalls Z-0189-2022
  5. Recall initiated FDA recalls Z-0192-2022
  6. Cleared FDA 510(k) K220295
  7. Listed FDA GUDID 04573596214510
  8. Registered EUDAMED 457359621C253A46

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States FDA 510(k) Active

Specification

EUDAMED · GUDID

Generic name
Extracorporeal ultrasound imaging transducer, hand-held
Model / reference
04573596214510 EUDAMED
C253A GUDID
Primary DI / UDI-DI
04573596214510
Basic UDI-DI
457359621C253A46
Direct marketing DI
04573596214510
FDA product code
ITX, IYO, IYN
EMDN code
Z110402010201 Curvilinear transcutaneous two–dimensional ultrasound probes (convex–array)
GMDN term
Extracorporeal ultrasound imaging transducer, hand-held
Device class
Class IIa EUDAMED
Class II GUDID
Regulation number
892.1570, 892.1560, 892.1550
EU registration type
Standard device
EU market of registration
Netherlands
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 3 fields differ

FieldEUDAMEDGUDID
Model / referenceEUDAMED
04573596214510
GUDID
C253A
Device classEUDAMED
Class IIa
GUDID
Class II
Sterilise before useEUDAMED
Yes
GUDID
No

Registry differences

EUDAMED and FDA GUDID describe the same product (same UDI-DI) and state different values:

  • Sterilise before use UDI-DI 04573596214510: EUDAMED states Yes, FDA GUDID states No.

Device class is stated on each registry's own scale: Class IIa in the EU (EUDAMED), Class II in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.

Manufacturer details

EUDAMED SRN
JP-MF-000010401
Authorised representative
FUJIFILM Healthcare Europe GmbH DE-AR-000040920

Similar devices in this category

Other devices filed under CURVILINEAR TRANSCUTANEOUS TWO–DIMENSIONAL ULTRASOUND PROBES (CONVEX–ARRAY).

Other devices from Fujifilm Corporation

See all devices from Fujifilm Corporation →

Properties

  • Sterilise before use YesEUDAMEDNoGUDID
  • Prescription use
  • Serial number
  • Manufacturing date
Other properties (21, all answered No)