Chiron Diagnostics Corporation Acs: Centaur Br Immunoassay
510(k) K982680
Active
510(k)
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK982680 | Date15 Oct 1998 | StatusActive |
Recalls
-
- Ongoing Z-2425-2021
- Ongoing Z-2426-2021
- Ongoing Z-2424-2021
Product does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days. IFU calibration updated interval from 10 days to 3 days for all current and future kit lot numbers
-
- Terminated Z-1748-2017
- Terminated Z-1747-2017
- Terminated Z-1749-2017
Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems
-
- Terminated Z-0507-2013
- Terminated Z-0512-2013
Operating range of the ADVIA Centaur systems is 18C to 30C (64.4F to 86.0F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.
-
- Terminated Z-0162-2013
BR Assay for CA 27.29 Calibration Interval Change due quality control (QC) results outside of the expected ranges.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MOI
- Device Class
- Class II
- Regulation
- 21 CFR Part 866.6010
- Advisory Committee
- Immunology
- FDA Decision
- SESE Substantially Equivalent