Chol CH3810
Ref CH3810
Chol CH3810 is a total cholesterol lipid IVD (EU IVDD general IVD; FDA Class I; FDA product code CHH, Enzymatic Esterase--Oxidase, Cholesterol) made by Randox Laboratories, Ltd., registered in EUDAMED on 28 Oct 2026 (Basic UDI-DI B-05055273201284), cleared under FDA 510(k) K923504 on 10 Jan 1994 and listed in FDA GUDID since 24 Dec 2018 (UDI-DI 05055273201284), active as of 7 Nov 2026. 2 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
None ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/IVDD EUDAMED | NumberB-05055273201284 05055273201284 | Date28 Oct 2026 | StatusActive |
| United States | GUDID listing GUDID | Number05055273201284 | Date24 Dec 2018 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK923504 | Date10 Jan 1994 | StatusActive |
Official records
- EUDAMED public site B-05055273201284
- AccessGUDID record 05055273201284
- FDA 510(k) database K923504
From the FDA record
- Device description
- For the quantitative in vitro determination of cholesterol in serum and plasma.
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI B-05055273201284 registered 28 Oct 2026 Active
- FDA 510(k) K923504 decided 10 Jan 1994 Active
- FDA GUDID UDI-DI 05055273201284 listed 24 Dec 2018 Listed
Certificates
No certifier certificate is listed. Devices of this class are self-declared by the manufacturer.
Recalls
-
- Terminated Z-1424-2022
Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range; Cholesterol (CHOL) Instrument Specific Application using the recommended calibration material CAL2351, the product fails to meet its linearity performance claim quoted on the kit inserts. The linearity is reduced by up to -35% however Quality Control results will still fall within assigned ranges. This may lead to delayed results.
-
- Terminated Z-0328-2022
When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range
-
- Terminated Z-1652-2016
Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific
Timeline
- Cleared FDA 510(k) K923504
- Recall initiated FDA recalls Z-1652-2016
- Listed FDA GUDID 05055273201284
- Recall initiated FDA recalls Z-0328-2022
- Recall initiated FDA recalls Z-1424-2022
- Registered EUDAMED B-05055273201284
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Cholesterol (Chol) EUDAMEDChol GUDID
- Generic name
- Total cholesterol lipid IVD, kit, spectrophotometry
- Model / Ref / Catalog
- CH3810
- Primary DI / UDI-DI
- 05055273201284
- Basic UDI-DI
- B-05055273201284
- 510(k) Number
- FDA product code
- CHH
- EMDN code
- W01010205 Cholesterol
- GMDN term
- Total cholesterol lipid IVD, kit, spectrophotometry
- Device class
- Class IVD General EUDAMEDClass I GUDID
- Regulation number
- 862.1175
- EU registration type
- Legacy device (under a directive)
- EU market of registration
- Northern Ireland
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class IVD General | Class I |
Similar devices in this category
Other devices filed under CHOLESTEROL.
- Cholesterol – Liquizyme CHOD-PAP (Single Reagent) 230 003 by Spectrum for Diagnostic Industries (Joint Stock Company)
- ARCHITECT 04S9230 by ABBOTT IRELAND
- small dense LDL Seiken 08834059190 by Denka Co., Ltd.
- Cholesterol CHOD-PAP D-613 002 by Kalians Corporation
- CREATINE KINASE MB ISOENZYME REAGENT KIT(IMMUNOINHIBITION) 3004586 by DIRUI INDUSTRIAL CO., LTD
- CHOLESTEROL REAGENT KIT(ENZYME METHOD) 232012502002 by DIRUI INDUSTRIAL CO., LTD
- TCH Reagent Kit-2 (CHOD-PAP Enzymatic-Colorimetric Method) CC1076 by Getein Biotech, Inc
- BIOBASE R1:64mL×2 R2:16mL×2 R1:64mL×2 R2:16mL×2 by Biobase Biodustry(Shandong) Co.,Ltd.
- BIOBASE R:20mL×8 by Biobase Biodustry(Shandong) Co.,Ltd.
- ARCHITECT 7D62-22 07D6222 by Abbott