Comprehensive® Instruments 110003489
Listed
Valid until
Not applicable
FDA recalls
1 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00887868462658 | Date4 Mar 2022 | StatusListed |
Recalls
-
- Ongoing Z-1403-2025
Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.
-
- Terminated Z-2216-2017
Zimmer Biomet is conducting a voluntary recall for a single lot of the Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System after it was determined that two pieces from the lot were moved to finished goods and shipped despite having an open Non-Conformance Report (NCR). The NCR indicated that one piece from the lot had scratches and the other non-confirming threads.
-
- Terminated Z-1859-2017
Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
-
- Terminated Z-1658-2008
The glenoid implant may not seat fully on the face of the natural glenoid due to inadequate measurement tolerances (tolerance stack-up) in the design of the cutter. Use of the instrument may lead to incomplete seating of the glenoid implant, elevating the theoretical risk of glenoid loosening, which may lead to early revision surgery. The problem may contribute to early device failure and patien
-
- Terminated Z-0660-2013
Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events: 1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is implanted, then a gap of 1-3 mm will be prese
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- Comprehensive® Instruments
- Generic Name
- Instrument tray, reusable
- Model / Ref / Catalog
- 110003489
- Primary DI / UDI-DI
- 00887868462658
- FDA Product Code
- KWT, MBF, HSD, KWS
- GMDN Term
- Instrument tray, reusable
- Device Class
- Class II
- Regulation Number
- 888.3650, 888.3670, 888.3690, 888.3660
- GUDID distribution status
- In Commercial Distribution
- Sterilization Method
- Moist Heat or Steam Sterilization
Other devices from Biomet Orthopedics, Inc.
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No