Comprehensive Shoulder System 118000
Comprehensive Shoulder System 118000 is a reverse shoulder prosthesis base plate (FDA Class II) made by Biomet Orthopedics, Inc., cleared under FDA 510(k) K202232 on 19 Feb 2021 (3 earlier submissions) and listed in FDA GUDID since 24 Oct 2015 (UDI-DI 00880304531048), active as of 23 Oct 2026. 20 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK202232 | Date19 Feb 2021 | StatusActive |
| United States | GUDID listing | Number00880304531048 | Date24 Oct 2015 | StatusListed |
Official records
- AccessGUDID record 00880304531048
- FDA 510(k) database K202232
- FDA 510(k) database K193373
- FDA 510(k) database K120121
- FDA 510(k) database K080642
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K202232 decided 19 Feb 2021 Active
- FDA GUDID UDI-DI 00880304531048 listed 24 Oct 2015 Listed
Recalls
-
- Terminated Z-0619-2021
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
-
- Terminated Z-0093-2021
Product potentially being packaged without a taper adapter or baseplate. The package should contain one (1) baseplate and one (1) taper adapter, product may be packaged with two (2) baseplates and no (0) adapters or with no (0) baseplates and two (2) adapters
-
- Terminated Z-2144-2020
Potential presence of elevated endotoxin levels that exceed the specification limit
-
- Terminated Z-1750-2020
There were unknown bioburden levels prior to sterilization, which may affect sterility of the product.
-
- Terminated Z-1615-2020
Package is intended to contain one (1) baseplate and one (1) taper adapter; however, devices may be packaged with two (2) baseplates and no (0) adapters.
-
- Terminated Z-0972-2019
The product was potentially being packaged without a taper adapter.
-
- Terminated Z-1859-2017
Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
-
- Terminated Z-1103-2017
Higher than anticipated rate of fracturing due to design. Risk of revision surgery due to fracturing.
-
- Terminated Z-1208-2015
The taper adaptor is missing from the packaging, which may result in a delay in surgery greater than 30 minutes.
-
- Terminated Z-1548-2013
Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction. When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone.
-
- Terminated Z-2548-2011
- Terminated Z-2547-2011
- Terminated Z-2550-2011
- Terminated Z-2546-2011
- Terminated Z-2542-2011
- Terminated Z-2545-2011
- Terminated Z-2541-2011
- Terminated Z-2549-2011
- Terminated Z-2544-2011
- Terminated Z-2543-2011
The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.
-
- Terminated Z-0712-2011
- Terminated Z-0713-2011
The firm received complaints regarding fracturing of the device.
Timeline
- Recall initiated FDA recalls Z-1103-2017
- Recall initiated FDA recalls Z-1859-2017
- Recall initiated FDA recalls Z-0972-2019
- Recall initiated FDA recalls Z-1615-2020
- Recall initiated FDA recalls Z-1750-2020
- Cleared FDA 510(k) K193373
- Recall initiated FDA recalls Z-2144-2020
- Recall initiated FDA recalls Z-0093-2021
- Recall initiated FDA recalls Z-0619-2021
- Cleared FDA 510(k) K202232
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Comprehensive Shoulder System
- Generic name
- Reverse shoulder prosthesis base plate
- Model / Ref / Catalog
- 118000
- Primary DI / UDI-DI
- 00880304531048
- FDA product code
- KWS, PHX
- GMDN term
- Reverse shoulder prosthesis base plate
- Device class
- Class II
- Regulation number
- 888.3660
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category
Other devices filed under Reverse shoulder prosthesis base plate.
- Glenoid Reconstruction System 04.01.0300 by Medacta International SA
- DELTA XTEND 130760000 by DEPUY (IRELAND)
- Glenoid Reconstruction System 04.01.0304 by Medacta International SA
- LINK Embrace Shoulder System 645-080/30 by WALDEMAR LINK GmbH & Co. KG
- DELTA XTEND 1407-60-010 140760010 by DEPUY (IRELAND)
- Anatomical Shoulder™ 01.04223.200 by Zimmer, Inc.