CRW® Crwma
CRW Crwma is a stereotactic radiosurgical invasive head frame (FDA Class II; FDA product code HAW, Neurological stereotaxic Instrument) made by Integra Lifesciences Corp., cleared under FDA 510(k) K944463 on 26 Oct 1995 and listed in FDA GUDID since 26 Jul 2016 (UDI-DI 10381780251507), active as of 23 Oct 2026. 4 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10381780251507 | Date26 Jul 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK944463 | Date26 Oct 1995 | StatusActive |
Official records
- AccessGUDID record 10381780251507
- FDA 510(k) database K944463
From the FDA record
- Device description
- MAYFIELD® Adaptor for CRW® System
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K944463 decided 26 Oct 1995 Active
- FDA GUDID UDI-DI 10381780251507 listed 26 Jul 2016 Listed
Recalls
-
- Terminated Z-0025-2016
Integra identified that the UCHRAP component (Arc Adapter Plate) would not assemble properly to the UCHRAR component (ARC Adapter Ring) of the Compact Head Ring Adapter due to an error in the manufacturing drawing.
-
- Terminated Z-2428-2012
- Terminated Z-2429-2012
T-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened, the T-handle could stop in a vertical position that prevents the CRWPRECISE and certain models of the CRW-ASL from seating properly on the head ring although it appears to be fully seated
-
- Terminated Z-1400-2011
Overall length of the intubation hoop in the HRAIM Intubation Head Ring Assembly is too long and will not allow a device to to attach
Timeline
- Cleared FDA 510(k) K944463
- Recall initiated FDA recalls Z-1400-2011
- Recall initiated FDA recalls Z-2428-2012
- Recall initiated FDA recalls Z-2429-2012
- Recall initiated FDA recalls Z-0025-2016
- Listed FDA GUDID 10381780251507
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Crw®
- Generic name
- Stereotactic radiosurgical invasive head frame
- Model / reference
- CRWMA
- Primary DI / UDI-DI
- 10381780251507
- 510(k) Number
- FDA product code
- HAW
- GMDN term
- Stereotactic radiosurgical invasive head frame
- Device class
- Class II
- Regulation number
- 882.4560
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code HAW (Neurological stereotaxic Instrument).
- THE MAYFIELD / ACCISS SYSTEM WITH WINDOWS NT FOR CRANIAL SURGERY, THE MAYFIELD / OPTICAL ACCISS SYSTEM WITH WINDOWS NT F by Ohio Medical Instrument Co., Inc.
- Alignment System Cranial, Alignment Software Cranial, Cirq Alignment Software Cranial Biopsy, Cirq Alignment Software Cranial sEEG, Varioguide Alignment Software Cranial by Brainlab AG
- Gouda Frame for Stereotactic Neurosurgery by Neurosurgical Devices, Inc.
- NAVIGATION SW HIP 3.1 ON CI by Brainlab AG
- Autoguide 9736380 by Medtronic Navigation Inc.
- Catheter Introducer for StealthStation System by Medtronic Surgical Navigation Technologies
- Stryker Navigation System, Stryker Fluoroscopy System, Virtual Fluoroscopy by Stryker Leibinger
- Brigit Surgical Device by Zimmer, Inc.
- 6007-004-100 by Stryker Leibinger
- SpineaSSIST System by Mazor Surgical Technologies, Ltd.