Skip to content

Cyberknife Robotic Radiosurgery System

by Accuray, Inc. 2 variants

510(k) K091999

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

3 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK091999 Date18 Sep 2009ClearedStatusActive

Recalls

3 recalls · FDA

    • Terminated Z-1098-2012 FDA Not yet classified initiated 16 Jan 2012

    An anomaly was discovered during internal regression testing, where the dose information is displayed incorrectly during treatment planning during a specific workflow using the optional 4D Ray Tracing dose calculation algorithm. As a result, it is possible that the dose calculation will display a lower dose than the intended dose prescribed for treatment delivery.

    • Terminated Z-0382-2012 FDA Unclassified (pre-2012) initiated 7 Nov 2011

    User Facility reported an anomaly where inhale and exhale CT pairs used for treatment planning did not represent the same magnitude of respiration that was displayed during treatment delivery.

    • Terminated Z-0714-2011 FDA Unclassified (pre-2012) initiated 28 Oct 2010

    If users modify the dose calculation box to cover just the contoured anatomy and not the entire CT volume, dosage may not be properly calculated.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
IYE
Device Class
Class II
Regulation
21 CFR Part 892.5050
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from Accuray, Inc.