Accuray, Inc.
United StatesFDA applicant, unverified
- Devices
- 10
- FDA recalls
- 13
At a glance
Market reach
- United States
- FDA recalls
Catalogue
- Regimes: 510(k) Registry: 510(k)
- Regimes: 510(k) Registry: 510(k)
- Regimes: 510(k) Registry: 510(k)
- Regimes: 510(k) Registry: 510(k)
Cyberknife Robotic Radiosurgery System Ref K062967
Regimes: 510(k) Registry: 510(k)- Regimes: 510(k) Registry: 510(k)
Cyberknife System Ref K011024
Regimes: 510(k) Registry: 510(k)Cybernife Stereotactic Radiosurgery System Ref K051941
Regimes: 510(k) Registry: 510(k)Cyberknife Robotic Radiosurgery System Ref K091999
Regimes: 510(k) Registry: 510(k)Motion Tracking Ref K020294
Regimes: 510(k) Registry: 510(k)
FDA recalls
2010
- 28 Oct 2010 Z-0714-2011 Unclassified (pre-2012) Terminated If users modify the dose calculation box to cover just the contoured anatomy and not the entire CT volume, dosage may not be properly calculated.
Full distribution text
- 6 Jul 2010 Z-0890-2011 Unclassified (pre-2012) Terminated Latches responsible for securing the cover to the device may come lose if not properly secured and the cover may unexpectedly drop off. A design change is planned.
Full distribution text
- 15 Jan 2010 Z-1023-2011 Unclassified (pre-2012) Terminated When a user modified the align center of an existing plan during the treatment planning process the effective depths of each beam in the plan were not updated by the system. These depths were incorrect and led to an incorrect dose calculation. Software patch /update to address this issue. Implementation at affected sites will initiate, once the patch is available.
Full distribution text
- 28 Oct 2010 Z-0714-2011 Unclassified (pre-2012) Terminated If users modify the dose calculation box to cover just the contoured anatomy and not the entire CT volume, dosage may not be properly calculated.
2009
- 29 Dec 2009 Z-1008-2011 Unclassified (pre-2012) Terminated If electron density values are left empty, the calculation of radiation dose in a patient will be modeled as air-like density material rather than correct density. A plan may be created and saved, thus creating the risk of mistreatment.
Full distribution text
- 21 Jul 2009 Z-0853-2011 Unclassified (pre-2012) Terminated Mechanical Failure-- Patient Support System's table top may unexpectedly descend several inches.
Full distribution text
- 22 Jun 2009 Z-1810-2009 Unclassified (pre-2012) Terminated Targeting accuracy out of specification, Error alert does not render the system down, which may result in mistreatment in the wrong area.
Full distribution text
- 16 Mar 2009 Z-0908-2011 Unclassified (pre-2012) Terminated The product has the potential for the extension coupling of the Synchrony extension mounting hardware to become loose over time. The assembly weighs approximately 17 pounds; an unexpected contact with users or patients may result in serious injury if the boom descends unexpectedly.
Full distribution text
- 16 Mar 2009 Z-0909-2011 Unclassified (pre-2012) Terminated The product has the potential for the extension coupling of the Synchrony extension mounting hardware to become loose over time. The assembly weighs approximately 17 pounds; an unexpected contact with users or patients may result in serious injury if the boom descends unexpectedly.
Full distribution text
- 29 Dec 2009 Z-1008-2011 Unclassified (pre-2012) Terminated If electron density values are left empty, the calculation of radiation dose in a patient will be modeled as air-like density material rather than correct density. A plan may be created and saved, thus creating the risk of mistreatment.
2008
- 14 Nov 2008 Z-0629-2009 Unclassified (pre-2012) Terminated System may use random incorrect data to calculate dose. Resulting dose calculation error can exceed 100% of correct dose which may lead to serious patient injury.
- 30 Oct 2008 Z-0856-2009 Unclassified (pre-2012) Terminated Product may not be tensioned properly, potentially causing unexpected rotation or descent.
Show all 13 recalls
2008
- 1 May 2008 Z-1862-2008 Unclassified (pre-2012) Terminated Couch may move unexpectedly, which may result in patient impacting the linear accelerator.
Full distribution text
- 9 Apr 2008 Z-2056-2008 Unclassified (pre-2012) Terminated Sample beam data (which should not be used to treat patients) may differ from actual radiation output of an installed product, and which may be used by users.
Full distribution text
- 1 May 2008 Z-1862-2008 Unclassified (pre-2012) Terminated Couch may move unexpectedly, which may result in patient impacting the linear accelerator.
2007
- 21 Mar 2007 Z-0364-2008 Unclassified (pre-2012) Terminated Software problem - Sagittal, oblique or coronal sliced MR Images imported from a non-Accuray manufactured device into the product may result in inverted image orientation.
Full distribution text
- 21 Mar 2007 Z-0364-2008 Unclassified (pre-2012) Terminated Software problem - Sagittal, oblique or coronal sliced MR Images imported from a non-Accuray manufactured device into the product may result in inverted image orientation.
Frequently asked questions
How many devices does Accuray, Inc. have in the registry?
The registry currently lists 10 devices for Accuray, Inc.. Devices recorded under merged duplicate company entries are counted once.
Which registries list Accuray, Inc.'s devices?
Among its 10 most recently registered devices, Accuray, Inc. has devices listed in FDA 510(k) (10 devices).
What risk classes do Accuray, Inc.'s devices fall under?
Among its 10 most recently registered devices, the classes recorded for Accuray, Inc.'s devices are Class II. Each device page shows its own class and authorisation.
Are there FDA recall records for Accuray, Inc.?
FDA recall records list 13 recalls initiated by Accuray, Inc., the most recent initiated on 28 Oct 2010. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.
What are the newest devices from Accuray, Inc. in the registry?
The most recently registered devices listed for Accuray, Inc. include THE CYBERKNIFE SYSTEM FOR STEREOTACTIC RADIOSURGERY/ RADIOTHERAPY V2.1, MODIFICATION TO CYBERKNIFE ROBOTIC RADIOSURGERY SYSTEM, CYBERKNIFE SYSTEM FOR STEROTACTIC RADIOSURGERY/RADIOTHERAPY and CYBERKNIFE SYSTEM FOR STEREOTACTIC RADIOSURGERY/RADIOTHERAPY.