Skip to content

Cytrellis Dermal Micro-Coring System

by Cytrellis Biosystems, Inc.

510(k) K202517

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK202517 Date9 Jul 2021ClearedStatusActive

Official records

From the FDA record

Received by the FDA
1 Sep 2020
FDA decision
9 Jul 2021 (Substantially Equivalent)
FDA review time
311 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K202517 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

2 recalls · 1 ongoing · FDA

    • Ongoing Z-2393-2023 FDA Class II initiated 23 Jun 2023

    Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring

    • Completed Z-1648-2023 FDA Class II initiated 7 Apr 2023

    Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.

Timeline

3 dated events

  1. Cleared FDA 510(k) K202517
  2. Recall initiated FDA recalls Z-1648-2023
  3. Recall initiated FDA recalls Z-2393-2023

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
QAI
Device class
Class II
Regulation
21 CFR Part 878.4430
Advisory committee
General, Plastic Surgery
FDA decision
SESE Substantially Equivalent

Overview

AI-generated summary, 1 Oct 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

The Cytrellis Dermal Micro-Coring System is a powered microneedle device designed for controlled removal of micro-cores of dermal and epidermal tissue. Classified as a Class II device under FDA regulations (878.4430), it is intended for use in aesthetic procedures to address moderate to severe wrinkles in the mid and lower facial regions. The system employs precise micro-coring technology to target sagging skin without invasive surgical intervention, facilitating tissue remodeling and skin tightening.

This device is supplied as a reusable system with single-use disposable components, intended for use in clinical settings by trained practitioners. It is not MRI-compatible and requires proper sterilization and maintenance as outlined in manufacturer instructions. The system is authorized for use in the United States following FDA 510(k) clearance (K202517) under the Subcommittee for Urology (SU) and General Plastic Surgery advisory committees. Multiple registration numbers (e.g., 3007645866) correspond to its various configurations and accessories, ensuring traceability and compliance with regulatory standards.

Devices cleared under K202517

25 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K202517
UDI-DISize and differences
Ellacor EL-03N01
CYTRELLIS BIOSYSTEMS, INC.
EL-03N01 / EL-03N01B933EL03N011
Single use: yes
On the market
Ellacor EL-03N01
CYTRELLIS BIOSYSTEMS, INC.
EL-03N01 / EL-03N01B933EL03N010
Single use: yes
On the market
Ellacor EL-03N02
CYTRELLIS BIOSYSTEMS, INC.
EL-03N02 / EL-03N02B933EL03N021
Single use: yes
On the market
Ellacor EL-03N02
CYTRELLIS BIOSYSTEMS, INC.
EL-03N02 / EL-03N02B933EL03N029
Single use: yes
On the market
Ellacor EL-03N02
CYTRELLIS BIOSYSTEMS, INC.
EL-03N02 / EL-03N02B933EL03N020
Single use: yes
On the market
Ellacor EL-CCV01
CYTRELLIS BIOSYSTEMS, INC.
EL-CCV01 / EL-CCV01B933ELCCV010
Single use: yes
On the market
Ellacor EL-CCV01
CYTRELLIS BIOSYSTEMS, INC.
EL-CCV01 / EL-CCV01B933ELCCV011
Single use: yes
On the market
ellacor EL-CLP01
CYTRELLIS BIOSYSTEMS, INC.
EL-CLP01 / EL-CLP01B933ELCLP011
Single use: no
On the market
Ellacor EL-CON01
CYTRELLIS BIOSYSTEMS, INC.
EL-CON01 / EL-CON01B933ELCON010
Single use: no
On the market
Ellacor EL-CON01-R
CYTRELLIS BIOSYSTEMS, INC.
EL-CON01-R / EL-CON01-RB933ELCON01R0
Single use: no
On the market
ellacor EL-CON02
CYTRELLIS BIOSYSTEMS, INC.
EL-CON02 / EL-CON02B933ELCON02
Single use: no
On the market
Ellacor EL-CON02-R
CYTRELLIS BIOSYSTEMS, INC.
EL-CON02-R / EL-CON02-RB933ELCON02R0
Single use: no
On the market
Ellacor EL-FTS01
CYTRELLIS BIOSYSTEMS, INC.
EL-FTS01 / EL-FTS01B933ELFTS010
Single use: no
On the market
Ellacor EL-FTS01-R
CYTRELLIS BIOSYSTEMS, INC.
EL-FTS01-R / EL-FTS01-RB933ELFTS01R0
Single use: no
On the market
Ellacor EL-HPC01
CYTRELLIS BIOSYSTEMS, INC.
EL-HPC01 / EL-HPC01B933ELHPC010
Single use: no
On the market
Ellacor EL-HPC01-R
CYTRELLIS BIOSYSTEMS, INC.
EL-HPC01-R / EL-HPC01-RB933ELHPC01R0
Single use: no
On the market
Ellacor EL-HPC02
CYTRELLIS BIOSYSTEMS, INC.
EL-HPC02 / EL-HPC02B933ELHPC02
Single use: no
On the market
Ellacor EL-HPC02-R
CYTRELLIS BIOSYSTEMS, INC.
EL-HPC02-R / EL-HPC02-RB933ELHPC02R0
Single use: no
On the market
ellacor EL-KIT01
CYTRELLIS BIOSYSTEMS, INC.
EL-KIT01 / EL-KIT01B933ELKIT010
Single use: yes
On the market
ellacor EL-KIT02
CYTRELLIS BIOSYSTEMS, INC.
EL-KIT02 / EL-KIT02B933ELKIT02
Single use: yes
On the market
ellacor EL-KIT03
CYTRELLIS BIOSYSTEMS, INC.
EL-KIT03 / EL-KIT03B933ELKIT03
Single use: yes
On the market
ellacor EL-KIT04
CYTRELLIS BIOSYSTEMS, INC.
EL-KIT04 / EL-KIT04B933ELKIT04
Single use: yes
On the market
ellacor EL-SYS01
CYTRELLIS BIOSYSTEMS, INC.
EL-SYS01 / EL-SYS01B933ELSYS01
Single use: no
On the market
ellacor EL-SYS02
CYTRELLIS BIOSYSTEMS, INC.
EL-SYS02 / EL-SYS02B933ELSYS02
Single use: no
On the market
ellacor EL-SYS02-R
CYTRELLIS BIOSYSTEMS, INC.
EL-SYS02-R / EL-SYS02-RB933ELSYS02R0
Single use: no
On the market

Similar devices

Other devices filed under FDA product code QAI (Powered Microneedle Device).

Other devices from Cytrellis Biosystems, Inc.

  • ellacor EL-KIT04 Open recall

    Class II
  • ellacor EL-SYS02-R Open recall

    Class II
  • ellacor EL-SYS01 Open recall

    Class II
  • ellacor EL-CON02 Open recall

    Class II

See all devices from Cytrellis Biosystems, Inc. →