Skip to content

ellacor EL-KIT04

by Cytrellis Biosystems, Inc.

ellacor EL-KIT04 is a cosmetic micro-needling electronic handpiece (FDA Class II; FDA product code QAI, Powered Microneedle Device) made by Cytrellis Biosystems, Inc., cleared under FDA 510(k) K202517 on 9 Jul 2021 and listed in FDA GUDID since 16 Dec 2025 (UDI-DI B933ELKIT04), active as of 23 Oct 2026. 2 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

510(k) K202517

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing NumberB933ELKIT04 UDI-DI Date16 Dec 2025ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK202517 Date9 Jul 2021DecidedStatusActive

Official records

From the FDA record

Device description
6 pack Consumable Kit (Triple Needle Cartridge,V2) includes the following items: Needle Cartridge (EL-03N02) Biofilter Suction Tube Cleaning Cover (EL-CCV01) Tube Clip (EL-CLP01)
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K202517 decided 9 Jul 2021 Active
  • FDA GUDID UDI-DI B933ELKIT04 listed 16 Dec 2025 Listed

Recalls

2 recalls · 1 ongoing · FDA

    • Ongoing Z-2393-2023 FDA Class II initiated 23 Jun 2023

    Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring

    • Completed Z-1648-2023 FDA Class II initiated 7 Apr 2023

    Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.

Timeline

4 dated events

  1. Cleared FDA 510(k) K202517
  2. Recall initiated FDA recalls Z-1648-2023
  3. Recall initiated FDA recalls Z-2393-2023
  4. Listed FDA GUDID B933ELKIT04

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
ellacor
Generic name
Cosmetic micro-needling electronic handpiece, professional
Model / Ref / Catalog
EL-KIT04
Primary DI / UDI-DI
B933ELKIT04
510(k) Number
FDA product code
QAI
GMDN term
Cosmetic micro-needling electronic handpiece, professional
Device class
Class II
Regulation number
878.4430
GUDID distribution status
In Commercial Distribution

Similar devices

Other devices filed under FDA product code QAI (Powered Microneedle Device).

Other devices from Cytrellis Biosystems, Inc.

  • ellacor EL-SYS02-R Open recall

    Class II
  • ellacor EL-SYS01 Open recall

    Class II
  • ellacor EL-CON02 Open recall

    Class II
  • Ellacor EL-HPC01 Open recall

    Class II

See all devices from Cytrellis Biosystems, Inc. →

Properties

  • Single use
  • Prescription use
  • Lot / batch number
  • Manufacturing date