Dade Indirect Imt System with TCO2 Electrode
by Dade Intl., Inc. 510(k)
Active
510(k) K970330
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK970330 | Date14 Apr 1997 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1398-2015
A positive bias beyond 5% resulting in a dilution check failure or an unnecessary dilution check correction if the positive bias is greater than 1% and less than 5%. The dilution check correction would be detected by QC; however, if not detected by QC, the calibration curve would be decreased and cause elevated sodium, potassium and chloride patient results.
Specification
- 510(k) Number
- FDA Product Code
- CEM
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1600
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent