daVinci 5 380730
Ref 380730
daVinci 5 380730 is a robotic surgical telemanipulation system (EU MDR Class IIb; FDA Class II; FDA product code NAY, System, surgical, computer controlled instrument) made by Intuitive Surgical, Inc., registered in EUDAMED on 28 Aug 2026 (Basic UDI-DI 08868741209899D), cleared under FDA 510(k) K232610 on 14 Mar 2024 and listed in FDA GUDID since 21 Mar 2024 (UDI-DI 00886874119747), active as of 25 Oct 2026. 6 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
6 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number08868741209899D 00886874119747 | Date28 Aug 2026 | StatusActive |
| United States | GUDID listing GUDID | Number00886874119747 | Date21 Mar 2024 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK232610 | Date14 Mar 2024 | StatusActive |
Official records
- EUDAMED public site 08868741209899D
- AccessGUDID record 00886874119747
- FDA 510(k) database K232610
From the FDA record
- Device description
- Surgeon Side Cart
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI 08868741209899D registered 28 Aug 2026 Active
- FDA 510(k) K232610 decided 14 Mar 2024 Active
- FDA GUDID UDI-DI 00886874119747 listed 21 Mar 2024 Listed
Recalls
-
- Ongoing Z-2310-2026
Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic endoscope.
-
- Ongoing Z-1355-2026
Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.
-
- Ongoing Z-1428-2026
An error event following the release of a software version for the surgical system that can result in loss of user interface content on an external monitor or tower monitor.
-
- Ongoing Z-1232-2026
Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion axis ball screws that exhibit the symptoms of being bent and misaligned which is a precursor to potential ball screw fatigue failure/fraction
-
- Ongoing Z-1325-2025
Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
-
- Ongoing Z-0964-2026
Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure
Timeline
- Cleared FDA 510(k) K232610
- Listed FDA GUDID 00886874119747
- Recall initiated FDA recalls Z-0964-2026
- Recall initiated FDA recalls Z-1325-2025
- Recall initiated FDA recalls Z-1232-2026
- Recall initiated FDA recalls Z-1428-2026
- Recall initiated FDA recalls Z-1355-2026
- Recall initiated FDA recalls Z-2310-2026
- Registered EUDAMED 08868741209899D
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- da Vinci 5 Console – Model: C5000 EUDAMEDdaVinci 5 GUDID
- Generic name
- Robotic surgical telemanipulation system
- Device name
- da Vinci IS5000 Surgical System - Console
- Model / reference
- 380730
- Primary DI / UDI-DI
- 00886874119747
- Basic UDI-DI
- 08868741209899D
- FDA product code
- NAY
- EMDN code
- Z12020101 Robot-assisted endoscopic surgery systems
- GMDN term
- Robotic surgical telemanipulation system
- Device class
- Class IIb EUDAMEDClass II GUDID
- Regulation number
- 876.1500
- EU registration type
- Standard device
- EU market of registration
- Germany
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class IIb | Class II |
Similar devices in this category
Other devices filed under ROBOT-ASSISTED ENDOSCOPIC SURGERY SYSTEMS.
- Dexter Handle Grip Dexter L6 146 by Distalmotion SA
- Sentire Surgical System C1002 by Cornerstone Technology (Shenzhen) Limited
- MULTIMED EM - by ELMED Elektronik ve Medikal San. Tic. A.S.
- Hugo™ MRASC0001 by Covidien llc
- Sentire Surgical System C1003 by Cornerstone Technology (Shenzhen) Limited
- Versius Camera Head with NIR functionality 90911 by CMR SURGICAL LIMITED
- Versius Endoscope 0 Degree with NIR Functionality 90910 by CMR SURGICAL LIMITED
- Dexter 487 by Distalmotion SA
- ILY 2003 by STERLAB