Force Feedback 475093
by Intuitive Surgical, Inc. 2 variants
Force Feedback 475093 is a robotic surgical forceps (FDA Class II; FDA product code NAY, System, surgical, computer controlled instrument) made by Intuitive Surgical, Inc., cleared under FDA 510(k) K232610 on 14 Mar 2024 and listed in FDA GUDID since 21 Mar 2024 (UDI-DI 00886874120729), active as of 22 Oct 2026. 6 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
6 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00886874120729 | Date21 Mar 2024 | StatusListed |
| United States | FDA 510(k)SESE | NumberK232610 | Date14 Mar 2024 | StatusActive |
Official records
- AccessGUDID record 00886874120729
- FDA 510(k) database K232610
From the FDA record
- Device description
- Force Feedback ProGrasp Forceps
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- FDA 510(k) K232610 decided 14 Mar 2024 Active
- FDA GUDID UDI-DI 00886874120729 listed 21 Mar 2024 Listed
Recalls
-
- Ongoing Z-2310-2026
Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic endoscope.
-
- Ongoing Z-1355-2026
Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.
-
- Ongoing Z-1428-2026
An error event following the release of a software version for the surgical system that can result in loss of user interface content on an external monitor or tower monitor.
-
- Ongoing Z-1232-2026
Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion axis ball screws that exhibit the symptoms of being bent and misaligned which is a precursor to potential ball screw fatigue failure/fraction
-
- Ongoing Z-1325-2025
Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
-
- Ongoing Z-0964-2026
Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure
Timeline
- Cleared FDA 510(k) K232610
- Listed FDA GUDID 00886874120729
- Recall initiated FDA recalls Z-0964-2026
- Recall initiated FDA recalls Z-1325-2025
- Recall initiated FDA recalls Z-1232-2026
- Recall initiated FDA recalls Z-1428-2026
- Recall initiated FDA recalls Z-1355-2026
- Recall initiated FDA recalls Z-2310-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Force Feedback
- Generic name
- Robotic surgical forceps, reusable
- Model / reference
- 475093
- Primary DI / UDI-DI
- 00886874120729
- FDA product code
- NAY
- GMDN term
- Robotic surgical forceps, reusable
- Device class
- Class II
- Regulation number
- 876.1500
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
- Size
- Outer Diameter — 8 Millimeter
Similar devices in this category
Other devices filed under Robotic surgical forceps, reusable.