Discovery Model Nm/CT 670
by Ge Medical Systems F.I. Haifa
510(k) K093514
Active
Valid until
No expiry
FDA recalls
6 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK093514 | Date10 Dec 2009 | StatusActive |
Recalls
-
- Ongoing Z-0895-2025
- Ongoing Z-0894-2025
- Ongoing Z-0893-2025
- Ongoing Z-0891-2025
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
-
- Ongoing Z-1053-2023
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
-
- Ongoing Z-1547-2021
GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific workflows. The issue occurs only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would expose the patient to additional X-ray radiation.
-
- Terminated Z-0857-2021
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
-
- Terminated Z-2998-2020
- Terminated Z-2996-2020
- Terminated Z-2997-2020
Rotor bearing screws were found loose on detector.
-
- Terminated Z-1260-2019
- Terminated Z-1259-2019
- Terminated Z-1265-2019
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
-
- Completed Z-0606-2019
- Completed Z-0607-2019
- Completed Z-0608-2019
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
-
- Terminated Z-2194-2017
It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.
-
- Terminated Z-0367-2014
GE Healthcare has recently become aware of a potential safety issue related to a SPECT scan procedure used for pulmonary imaging. Contact between the Detector (GE Healthcare Discovery NM/CT 670, and/or Optima NM/CT 640, and/or Discovery NM630, and/or Brivo NM 615 all configurations) and the Patient's elbow may occur during SPECT protocols using 450 angular steps in scenarios when a Patients a
-
- Terminated Z-1792-2013
GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
-
- Terminated Z-0121-2013
GE Healthcare has recently become aware of a potential safety issue associated with the table pallet of your Discovery NM/CT 670 and/or Discovery NM630 system which may impact operation safety. At scan setup, during gantry rotation towards scanning position, a potential contact between the table-pallet-edge and the detector was identified. Such occurrence may damage the pallet front-edge and
-
- Terminated Z-1110-2012
GE Healthcare has recently become aware of a potential safety issue associated with the collimator cart and lead cover of your Discovery NM/CT 670 and/or Discovery NM 630 system that may impact operator safety. 1. The cart locking pin release and cart disengagement could result in the collimator falling down and could lead to operator or service personnel injury. 2. Another issue is that the
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KPS
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1200
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent