GE Quasar Nuclear Medicine System

by Ge Medical Systems F.I. Haifa

510(k) K022960

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

5 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK022960 Date19 Sep 2002ClearedStatusActive

Recalls

24 recalls · 5 ongoing · FDA

    • Ongoing Z-2097-2025 FDA Class II initiated 20 Jun 2025

    Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.

  • 4 recalls, same reason

    • Ongoing Z-0888-2025 FDA Class II initiated 18 Dec 2024
    • Ongoing Z-0897-2025 FDA Class II initiated 18 Dec 2024
    • Ongoing Z-0899-2025 FDA Class II initiated 18 Dec 2024
    • Ongoing Z-0898-2025 FDA Class II initiated 18 Dec 2024

    GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

  • 5 recalls, same reason

    • Terminated Z-0186-2015 FDA Class II initiated 14 Oct 2014
    • Terminated Z-0184-2015 FDA Class II initiated 14 Oct 2014
    • Terminated Z-0189-2015 FDA Class II initiated 14 Oct 2014
    • Terminated Z-0188-2015 FDA Class II initiated 14 Oct 2014
    • Terminated Z-0185-2015 FDA Class II initiated 14 Oct 2014

    GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.

    • Terminated Z-0187-2015 FDA Class II initiated 14 Oct 2014

    GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems. This check is being added to prevent a portion of the system falling onto the patient during a scan due to fasteners being loose that secure the camera to the gantry.

  • 3 recalls, same reason

    • Terminated Z-1200-2014 FDA Class II initiated 7 Feb 2014
    • Terminated Z-1199-2014 FDA Class II initiated 7 Feb 2014
    • Terminated Z-1198-2014 FDA Class II initiated 7 Feb 2014

    Potential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on a single system, in which the detector moved to the home position and then slowly starting drifting inward toward the patient. This generated both a visual and audio alert condition on both the console and the gantry. N

    • Terminated Z-1787-2013 FDA Class I initiated 13 Jun 2013

    GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.

    • Terminated Z-1078-2013 FDA Class II initiated 6 Mar 2012

    Accelerated fatigue of the lateral axis motion subsystem of the Infinia Computed Emission Tomography System could result in mechanical failure causing the detector to slide; may impact patient and operators.

  • 2 recalls, same reason

    • Terminated Z-1095-2009 FDA Unclassified (pre-2012) initiated 8 Sep 2008
    • Terminated Z-1094-2009 FDA Unclassified (pre-2012) initiated 8 Sep 2008

    There is a possibility for misregistration when brain scans are performed with Infinia with the Hawkeye option. This misregistration could present the potential for misdiagnosis when the patient headrest sags under the weight of a head.

  • 3 recalls, same reason

    • Terminated Z-1120-2009 FDA Unclassified (pre-2012) initiated 2 Jun 2008
    • Terminated Z-1121-2009 FDA Unclassified (pre-2012) initiated 2 Jun 2008
    • Terminated Z-1119-2009 FDA Unclassified (pre-2012) initiated 2 Jun 2008

    The metal pin in the adjustable head holder, which is used to acquire Brain SPECT data on Infinia systems (Infinia, Infinia Hawkeye, and Infinia Hawkeye 4), creates an artifact on the patient's cerebellum, which may lead to the potential for misdiagnosis.

  • 3 recalls, same reason

    • Terminated Z-1002-2009 FDA Unclassified (pre-2012) initiated 2 Jun 2008
    • Terminated Z-1001-2009 FDA Unclassified (pre-2012) initiated 2 Jun 2008
    • Terminated Z-1003-2009 FDA Unclassified (pre-2012) initiated 2 Jun 2008

    GE Healthcare has recently became aware of excessive pressure applied by the collimator pressure sensitive cover of Infinia (Infinia, Infinia Hawkeye, Infinia Hawkeye 4) that may impact patient safety. It was reported at a customer site that the pressure used to activate the Pressure Sensitive Device (PSD) should be lowered and that the body contouring software requires correction to prevent po

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
KPS
Device Class
Class II
Regulation
21 CFR Part 892.1200
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from Ge Medical Systems F.I. Haifa

See all devices from Ge Medical Systems F.I. Haifa →