GE Quasar Nuclear Medicine System
by Ge Medical Systems F.I. Haifa
510(k) K022960
Active
Valid until
No expiry
FDA recalls
5 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK022960 | Date19 Sep 2002 | StatusActive |
Recalls
-
- Ongoing Z-2097-2025
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
-
- Ongoing Z-0888-2025
- Ongoing Z-0897-2025
- Ongoing Z-0899-2025
- Ongoing Z-0898-2025
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
-
- Terminated Z-0186-2015
- Terminated Z-0184-2015
- Terminated Z-0189-2015
- Terminated Z-0188-2015
- Terminated Z-0185-2015
GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.
-
- Terminated Z-0187-2015
GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems. This check is being added to prevent a portion of the system falling onto the patient during a scan due to fasteners being loose that secure the camera to the gantry.
-
- Terminated Z-1200-2014
- Terminated Z-1199-2014
- Terminated Z-1198-2014
Potential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on a single system, in which the detector moved to the home position and then slowly starting drifting inward toward the patient. This generated both a visual and audio alert condition on both the console and the gantry. N
-
- Terminated Z-1787-2013
GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.
-
- Terminated Z-1078-2013
Accelerated fatigue of the lateral axis motion subsystem of the Infinia Computed Emission Tomography System could result in mechanical failure causing the detector to slide; may impact patient and operators.
-
- Terminated Z-1095-2009
- Terminated Z-1094-2009
There is a possibility for misregistration when brain scans are performed with Infinia with the Hawkeye option. This misregistration could present the potential for misdiagnosis when the patient headrest sags under the weight of a head.
-
- Terminated Z-1120-2009
- Terminated Z-1121-2009
- Terminated Z-1119-2009
The metal pin in the adjustable head holder, which is used to acquire Brain SPECT data on Infinia systems (Infinia, Infinia Hawkeye, and Infinia Hawkeye 4), creates an artifact on the patient's cerebellum, which may lead to the potential for misdiagnosis.
-
- Terminated Z-1002-2009
- Terminated Z-1001-2009
- Terminated Z-1003-2009
GE Healthcare has recently became aware of excessive pressure applied by the collimator pressure sensitive cover of Infinia (Infinia, Infinia Hawkeye, Infinia Hawkeye 4) that may impact patient safety. It was reported at a customer site that the pressure used to activate the Pressure Sensitive Device (PSD) should be lowered and that the body contouring software requires correction to prevent po
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KPS
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1200
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent