Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750W 3.0T, Optima MR450W 1.5T

by Ge Healthcare (Ge Medical Systems, LLC)

510(k) K132376

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

6 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK132376 Date15 Nov 2013ClearedStatusActive

Recalls

15 recalls · 6 ongoing · FDA

    • Ongoing Z-2302-2025 FDA Class II initiated 20 Jun 2025

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    • Ongoing Z-2468-2024 FDA Class II initiated 14 Jun 2024

    GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.

  • 4 recalls, same reason

    • Ongoing Z-0593-2022 FDA Class II initiated 23 Dec 2021
    • Ongoing Z-0602-2022 FDA Class II initiated 23 Dec 2021
    • Ongoing Z-0594-2022 FDA Class II initiated 23 Dec 2021
    • Ongoing Z-0601-2022 FDA Class II initiated 23 Dec 2021

    The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.

  • 4 recalls, same reason

    • Terminated Z-3023-2017 FDA Class II initiated 11 May 2017
    • Terminated Z-3022-2017 FDA Class II initiated 11 May 2017
    • Terminated Z-3024-2017 FDA Class II initiated 11 May 2017
    • Terminated Z-3025-2017 FDA Class II initiated 11 May 2017

    Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

    • Terminated Z-1188-2017 FDA Class II initiated 20 Jan 2017

    Potential safety issue with the patient bore heating on the Discovery MR450 narrow bore 1.5T MRI. In the event of a RF component failure, typically happening after extended system use, a small area on the bore surface may become warmer than normal. If this happens in an area where there is patient contact with the bore, and proper padding is not used, it may cause a serious burn.

    • Terminated Z-2131-2016 FDA Class II initiated 8 Jun 2016

    GE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w. A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used. No injury has been reported to GE as a result of this issue.

    • Terminated Z-0201-2019 FDA Class II initiated 7 Dec 2015

    GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue. UPDATE 5/27/2016: Recall has been updated to include an additional product which includes the Discovery 3.0T MR750 systems running Application Software DV24 R01, DV25 R02, or DV25.1 M3.

  • 2 recalls, same reason

    • Terminated Z-0254-2016 FDA Class II initiated 31 Aug 2015
    • Terminated Z-0259-2016 FDA Class II initiated 31 Aug 2015

    When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LNH
Device Class
Class II
Regulation
21 CFR Part 892.1000
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from Ge Healthcare (Ge Medical Systems, LLC)

  • SIGNA Creator, SIGNA Explorer Open recall

    Class II
  • Optima MR450W

    Class II
  • SIGNA Architect Open recall

    Class II
  • SIGNA Hero Open recall

    Class II
  • Ready View

    Class II
  • SIGNA Premier Open recall

    Class II

See all devices from Ge Healthcare (Ge Medical Systems, LLC) →