SIGNA Creator, SIGNA Explorer
by Ge Healthcare (Ge Medical Systems, LLC)
510(k) K143251
Active
Valid until
No expiry
FDA recalls
11 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK143251 | Date4 Feb 2015 | StatusActive |
Recalls
-
- Ongoing Z-1242-2025
- Ongoing Z-1239-2025
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
-
- Ongoing Z-2473-2024
- Ongoing Z-2475-2024
- Ongoing Z-2472-2024
- Ongoing Z-2470-2024
- Ongoing Z-2476-2024
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
- Ongoing Z-0596-2022
- Ongoing Z-0597-2022
The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
-
- Ongoing Z-2058-2021
- Ongoing Z-2059-2021
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.
-
- Terminated Z-0137-2022
Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
-
- Terminated Z-3026-2017
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
-
- Terminated Z-2731-2016
GE Healthcare is conducting a recall dealing with an incorrect RF Power Monitor Function involving SIGNA CREATOR and SIGNA EXPLORER which could result in higher than expected thermal dose to the patient and higher than expected localized heating. There have been no failures of this type reported, and no injuries reported as a result of this issue.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LNH
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent