SIGNA Creator, SIGNA Explorer

by Ge Healthcare (Ge Medical Systems, LLC)

510(k) K143251

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

11 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK143251 Date4 Feb 2015ClearedStatusActive

Recalls

14 recalls · 11 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-1242-2025 FDA Class II initiated 31 Jan 2025
    • Ongoing Z-1239-2025 FDA Class II initiated 31 Jan 2025

    For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

  • 5 recalls, same reason

    • Ongoing Z-2473-2024 FDA Class II initiated 14 Jun 2024
    • Ongoing Z-2475-2024 FDA Class II initiated 14 Jun 2024
    • Ongoing Z-2472-2024 FDA Class II initiated 14 Jun 2024
    • Ongoing Z-2470-2024 FDA Class II initiated 14 Jun 2024
    • Ongoing Z-2476-2024 FDA Class II initiated 14 Jun 2024

    GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.

  • 2 recalls, same reason

    • Ongoing Z-0596-2022 FDA Class II initiated 23 Dec 2021
    • Ongoing Z-0597-2022 FDA Class II initiated 23 Dec 2021

    The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.

  • 2 recalls, same reason

    • Ongoing Z-2058-2021 FDA Class II initiated 4 Jun 2021
    • Ongoing Z-2059-2021 FDA Class II initiated 4 Jun 2021

    GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.

    • Terminated Z-0137-2022 FDA Class II initiated 26 Aug 2021

    Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.

    • Terminated Z-3026-2017 FDA Class II initiated 11 May 2017

    Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

    • Terminated Z-2731-2016 FDA Class II initiated 31 Aug 2016

    GE Healthcare is conducting a recall dealing with an incorrect RF Power Monitor Function involving SIGNA CREATOR and SIGNA EXPLORER which could result in higher than expected thermal dose to the patient and higher than expected localized heating. There have been no failures of this type reported, and no injuries reported as a result of this issue.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LNH
Device Class
Class II
Regulation
21 CFR Part 892.1000
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent