Djo Surgical 508-03-436
Djo Surgical 508-03-436 is a device of GMDN type total reverse shoulder prosthesis (FDA Class II) made by Encore Medical LP, cleared under FDA 510(k) K172351 on 30 Oct 2017 and listed in FDA GUDID since 13 Dec 2017 (UDI-DI 00190446256977), active as of 8 Nov 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00190446256977 | Date13 Dec 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK172351 | Date30 Oct 2017 | StatusActive |
Official records
- AccessGUDID record 00190446256977
- FDA 510(k) database K172351
From the FDA record
- Device description
- ALTIVATE REVERSE,SMALL SOCKET INSERT,36MM SEMI CONSTRAINED+4
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- Labelled as not made with natural rubber latex
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K172351 decided 30 Oct 2017 Active
- FDA GUDID UDI-DI 00190446256977 listed 13 Dec 2017 Listed
Recalls
-
- Ongoing Z-0281-2024
Two different humeral socket Insert devices were swapped during packaging, which could result in surgeons not having the correct device for surgery, which could cause a delay in surgery.
-
- Ongoing Z-2607-2023
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
-
- Terminated Z-0117-2021
Product mislabeled. Use may result in a delay in surgery, stopping of surgery, and revision surgery.
Timeline
- Cleared FDA 510(k) K172351
- Listed FDA GUDID 00190446256977
- Recall initiated FDA recalls Z-0117-2021
- Recall initiated FDA recalls Z-2607-2023
- Recall initiated FDA recalls Z-0281-2024
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Djo Surgical
- Generic name
- Total reverse shoulder prosthesis
- Model / reference
- 508-03-436
- Primary DI / UDI-DI
- 00190446256977
- 510(k) Number
- FDA product code
- KWS, PHX, HSD
- GMDN term
- Total reverse shoulder prosthesis
- Device class
- Class II
- Regulation number
- 888.3660, 888.3690
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code KWS (PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED).
- SMR TT Hybrid Glenoid, SMR 3-Pegs Cemented Glenoid by Limacorporate
- Trabecular Metal® 00-4309-027-00 by Zimmer, Inc.
- AETOS Shoulder System by Smith & Nephew, Inc.
- Anatomical Shoulder™ 01.04211.073 by Zimmer, Inc.
- BUECHEL-PAPPAS HUMERAL STEM COMPONENT by ENDOTEC, INC.
- Shoulder - TiNbN Coating 04.01.0264 by Medacta International SA
- SMR TT Hybrid Glenoid 1379.59.103 by Limacorporate
- Anatomic Shoulder Prosthesis 04.01.0097 by Medacta International SA
- Comprehensive® Instrumentation 110029132 by Biomet Orthopedics, Inc.
- LINK Embrace Shoulder System 647-044/16 by WALDEMAR LINK GmbH & Co. KG