Endoskeleton TCS 5360-1405
Endoskeleton TCS 5360-1405 is a Class II device made by Titan Spine, Inc., cleared under FDA 510(k) K142940 on 28 Jan 2015 and listed in FDA GUDID since 17 Sep 2015 (UDI-DI M682536014050), active (FDA 510(k)), not in commercial distribution (GUDID) as of 23 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | NumberM682536014050 | Date17 Sep 2015 | StatusListed |
| United States | FDA 510(k)SESE | NumberK142940 | Date28 Jan 2015 | StatusActive |
Official records
- FDA 510(k) database K142940
Across registries
- FDA 510(k) K142940 decided 28 Jan 2015 Active
- FDA GUDID UDI-DI M682536014050 listed 17 Sep 2015 Listed
Recalls
-
- Terminated Z-2327-2015
Titan Spine is recalling specific Endoskeleton TCS Integrated Locking Bone Screw Assembly due to the locking bone screw collars not 'locking'.
-
- Terminated Z-1324-2015
The TCS Straight Hexlobular #6 Drivers (5210-1004) tip sheared off during use in procedures that involved patients with dense bones. As a result of this malfunction, Titan Spine, LLC is removing all of the TCS Straight Hex lobular #6 Drivers (5210-1004) Lot# G150101 from the field.
Timeline
- Cleared FDA 510(k) K142940
- Recall initiated FDA recalls Z-1324-2015
- Recall initiated FDA recalls Z-2327-2015
- Listed FDA GUDID M682536014050
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Endoskeleton TCS
- Generic name
- Metallic spinal interbody fusion cage
- Model / Ref / Catalog
- 5360-1405
- Primary DI / UDI-DI
- M682536014050
- 510(k) Number
- FDA product code
- ODP, OVE
- GMDN term
- Metallic spinal interbody fusion cage
- Device class
- Class II
- Regulation number
- 888.3080
- GUDID distribution status
- Not in Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
- Size
- Height — 5 MillimeterDepth — 14 MillimeterWidth — 16 Millimeter
Similar devices
Other devices filed under FDA product code ODP (Intervertebral fusion device with bone graft, cervical).
- ClariVy™ Cervical IBF System by VY SPINE LLC
- Tryptik Ti by Spineart
- DISCOVERY by Intuitive Spine, LLC
- Mecta-C Ti mark 03.28.306 by Medacta International SA
- Integral Titanium Cervical Interbody by NVISION BIOMEDICAL TECHNOLOGIES, INC
- CAPTIVA SPINE FUSELOX CERVICAL CAGE by Captiva Spine
- C.A. Medical Interbody Fusion System by C.A. Medical, LLC
- LnK Cervical Interbody Fusion Cage System by L&K Biomed
- Mecta-C Cervical Ti-Peek Cages 03.28.617 by Medacta International SA
- RIGEL PEEK Anterior Cervical Interbody Fusion System by MIRUS LLC