epoc Blood Analysis System
510(k) K200107
epoc Blood Analysis System is a medical device (FDA Class II; FDA product code CHL, Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph) made by Epocal Inc, cleared under FDA 510(k) K200107 on 5 Jun 2020, active as of 11 Oct 2026. 17 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
18 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK200107 | Date5 Jun 2020 | StatusActive |
Official records
- FDA 510(k) database K200107
From the FDA record
- Received by the FDA
- 17 Jan 2020
- FDA decision
- 5 Jun 2020 (Substantially Equivalent)
- FDA review time
- 140 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-2005-2026
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
-
- Ongoing Z-2272-2025
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.
-
- Ongoing Z-1685-2025
- Ongoing Z-1684-2025
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
-
- Ongoing Z-0741-2025
- Ongoing Z-0735-2025
- Ongoing Z-0739-2025
- Ongoing Z-0740-2025
- Ongoing Z-0734-2025
- Ongoing Z-0736-2025
- Ongoing Z-0738-2025
- Ongoing Z-0737-2025
Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
-
- Ongoing Z-0286-2025
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.
-
- Ongoing Z-0601-2024
Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under certain conditions, a software problem is causing erroneous results to be printed. When the results are run on the epoc NXS Host and Delete Blood Tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test. The worst possible outcome may result in a failure to diagnose hyperkalemia, due to an erroneous depressed potassium level or inappropriate treatment to hyperglycemia due to erroneous glucose reading.
-
- Ongoing Z-2477-2023
- Ongoing Z-2476-2023
- Ongoing Z-2478-2023
There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
-
- Ongoing Z-1215-2023
pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure
-
- Terminated Z-2970-2020
Sporadically inconsistent discrepant (low bias) glucose results on card lot 01-20095-10. Potential exists for a delay in diagnosis of hyperglycemia or unnecessary treatment for hypoglycemia.
Timeline
- Recall initiated FDA recalls Z-0739-2025
- Recall initiated FDA recalls Z-0740-2025
- Recall initiated FDA recalls Z-0734-2025
- Recall initiated FDA recalls Z-0736-2025
- Recall initiated FDA recalls Z-0738-2025
- Recall initiated FDA recalls Z-0737-2025
- Recall initiated FDA recalls Z-1685-2025
- Recall initiated FDA recalls Z-1684-2025
- Recall initiated FDA recalls Z-2272-2025
- Recall initiated FDA recalls Z-2005-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- CHL
- FDA applicant
- Epocal, Inc.
- Device class
- Class II
- Regulation
- 21 CFR Part 862.1120
- Advisory committee
- Clinical Chemistry
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K200107
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| epoc NXS Host 11413475 | 16266 / 11413475 | 00630414606095 | — | On the market |
Similar devices
Other devices filed under FDA product code CHL (Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph).
- NOVA STAT PROFILE PHOX BLOOD GAS ANALYZER by NOVA BIOMEDICAL CORPORATION
- NOVA STAT PROFILE 7 ANALYZER QUALITY CONTROL MATER by NOVA BIOMEDICAL CORPORATION
- ABL SYSTEM 620 by RADIOMETER AMERICA, INC.
- TB5 Ref Test/Blank Sensor 10327492 673396000 by Siemens Healthcare Diagnostics Inc.
- AVL BLOOD GAS INSTRUMENT MODEL 939 by JONES MEDICAL INSTRUMENT CO.
- 158 BLOOD GAS ANALYZER by Corning Medical & Scientific
- RAPIDPoint 500e Blood Gas System by Siemens Healthcare Diagnostics
- ABL90 FLEX by Radiometer Medical ApS
- COBAS B 123 POC SYSTEM, COBAS B 123 AUTOQC PACK TRI-LEVEL, COBAS B 123 AUTOCVC PACK, ROCHE COMBITROL PLUS B by Roche Diagnostics Corp.
- FLUSH SOLUTION CONCENTRATE-BLOOD GAS by Bionostics, Inc.